Transtracheal ventilation with a novel ejector-based device (Ventrain) in open, partly obstructed, or totally closed
M Paxian1, N P Preussler2, T Reinz2
1Department of Anaesthesiology and Intensive Care Medicine, Jena University Hospital, Erlanger Allee 101, 07747 Jena, Germany markus.paxian@med.uni-jena.de.
Background:
Transtracheal access and subsequent jet ventilation are among the last options in a 'cannot intubate-cannot oxygenate' scenario. These interventions may lead to hypercapnia, barotrauma, and haemodynamic failure in the event of an obstructed upper airway. The aim of the present study was to evaluate the efficacy and the haemodynamic effects of the Ventrain, a manually operated ventilation device that provides expiratory ventilation assistance. Transtracheal ventilation was carried out with the Ventrain in different airway scenarios in live pigs, and its performance was compared with a conventional jet ventilator.
Methods:
Pigs with open, partly obstructed, or completely closed upper airways were transtracheally ventilated either with the Ventrain or by conventional jet ventilation. Airway pressures, haemodynamic parameters, and blood gases obtained in the different settings were compared.
Results:
Mean (SD) alveolar minute ventilation as reflected by arterial partial pressure of CO2 was superior with the Ventrain in partly obstructed airways after 6 min in comparison with traditional manual jet ventilation [4.7 (0.19) compared with 7.1 (0.37) kPa], and this was also the case in all simulated airway conditions. At the same time, peak airway pressures were significantly lower and haemodynamic parameters were altered to a lesser extent with the Ventrain.
Conclusions:
The results of this study suggest that the Ventrain device can ensure sufficient oxygenation and ventilation through a small-bore transtracheal catheter when the airway is open, partly obstructed, or completely closed. Minute ventilation and avoidance of high airway pressures were superior in comparison with traditional hand-triggered jet ventilation, particularly in the event of complete upper airway obstruction.
More Related Videos
13:26Basic Surgical Techniques in the Göttingen Minipig: Intubation, Bladder Catheterization, Femoral Vessel Catheterization, and Transcardial Perfusion
Published on: June 26, 2011
07:20Lavage-induced Surfactant Depletion in Pigs As a Model of the Acute Respiratory Distress Syndrome ARDS
Published on: September 7, 2016
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...