Comparative meta-analysis of balloon-expandable and self-expandable valves for transcatheter aortic valve replacement
Shikhar Agarwal1, Akhil Parashar2, Dharam J Kumbhani3
1Department of Cardiovascular Medicine, J2-3, Heart and Vascular Institute, Cleveland Clinic, 9500 Euclid Avenue, Cleveland, OH 44195, United States.
Background:
Balloon-expandable valves (BEVs) and self-expandable valves (SEVs) are two major types of valves utilized for transcatheter aortic valve replacement (TAVR). We conducted a meta-analysis of available studies to compare the safety and efficacy of these two valve types.
Methods:
Medline search was conducted using standard search terms to determine eligible studies. Primary outcomes of the meta-analysis included death and stroke at 30days and 1year. Pooled estimates of procedural outcomes were also compared between the valve types. Analysis was performed for entire cohort and separately for patients undergoing transfemoral TAVR (TF-TAVR).
Results:
Analysis of entire cohort revealed similar 30-day mortality in the SEV and BEV cohorts. There was no significant difference in the incidence of stroke at 30days between the two study groups. Both pooled comparisons demonstrated a significant heterogeneity with I(2)>50%, necessitating the use of random effect modeling. We observed a significantly higher incidence of new pacemaker implantation, aortic regurgitation≥2+ at 30days, valve embolization, and need for >1 valve following SEV implantation compared with BEV implantation. Analysis of TF-TAVR cohort showed higher 30day [IRR (95% CI): 1.34 (1.19-1.52)] but a similar 1-year mortality [IRR (95% CI): 1.07 (0.96-1.19)] for SEV compared to BEV implantation.
Conclusion:
Compared to BEV implantation, SEV implantation was associated with a similar risk of mortality and stroke at 30-day and 1-year follow-up duration. Analysis of the TF-TAVR cohort revealed a significantly higher mortality at 30days among patients undergoing SEV implantation, compared with BEV implantation. In addition, there was a significantly higher incidence of other adverse events noted above, following SEV implantation, compared with BEV implantation.