Related Experiment Video
Updated: Apr 8, 2026

Novel In Vivo Micro-Computed Tomography Imaging Techniques for Assessing the Progression of Non-Alcoholic Fatty Liver Disease
Published on: March 24, 2023
Focal liver lesions detection: Comparison of respiratory-triggering, triggering and tracking navigator and
Said El Bouchaibi1, Kenneth Coenegrachts2, Maria Antonietta Bali1
1Department of Radiology, Clinics of MRI, Hôpital Erasme Université Libre de Bruxelles, 808 Route de Lennik B, 1070 Bruxelles, Belgium.
Purpose:
To compare low b value (10s/mm(2)) spin-echo echo-planar (SE-EP) diffusion-weighted imaging (DWI) acquired with respiratory-triggering (RT), triggering and tracking navigator (TT), tracking only navigator (TRON) techniques for image quality and focal liver lesions (FLL) detection in non-cirrhotic patients.
Material And Methods:
This bi-centric study was approved by the institutional review boards; informed consent was obtained. Eighty-three patients were prospectively included and SE-EP-DWI with RT, TT and TRON techniques were performed. DWI sequences were randomized and independently analyzed by two readers. The qualitative evaluation was based on a 3-point score for axial artifacts (motion, ghost, susceptibility artifacts and distortion) and stair-step artifacts. Sensitivity of FLL detection was calculated for all lesions together and after lesion size stratification (≤ 10 mm, >10-20mm and >20mm). The standard of reference consisted of a retrospective reading of the conventional MRI, the three DWI sequences and by follow-up (12 months): a total of 409 FLL were detected. Data between sequences was compared with non-parametric tests. Cohen's kappa coefficient was used for inter-observer agreement.
Results:
Image quality was comparable for RT and TT. TRON showed statistically significantly more axial artifacts for the two readers (p<0.05). Stair-step artifacts were not statistically significantly different between DWI sequences. Overall sensitivities for RT, TT, TRON were 85%, 86%, 82% and 86%, 89% 83%, respectively, for readers 1 and 2. The inter-observer agreement was very good.
Conclusion:
Image quality was better for RT and TT compared to TRON. Overall sensitivities for FLL detection were comparable between techniques and readers.