Can FLAIR hyperintense vessel (FHV) signs be influenced by varying MR parameters and flow velocities? A flow phantom
Sung Jun Ahn1, Kyung-Yul Lee2, Sung Soo Ahn1
1Department of Radiology, Yonsei University College of Medicine, Seoul, Republic of Korea.
Background:
Fluid-attenuated inversion recovery (FLAIR) hyperintense vessels (FHVs) have been used to assess leptomeningeal collateral flow in acute ischemic stroke. However, prior FHVs studies showed inconsistent results, which may be ascribable to different magnetic resonance (MR) parameters used.
Purpose:
To evaluate whether FHVs could be influenced by varying MR parameters and flow velocities, using a flow phantom.
Material And Methods:
A total of 512 sets of FLAIRs were performed with varying parameters and flow velocities, using a flow phantom. Flow phantom was manufactured with 3.5% agarose solution, an 8-mm inner diameter silicone tube and non-pulsatile pump. Varying MR parameters were repetition time (TR)/inversion time (TI), echo time (TE), flip angle (FA) of refocusing pulse, and periodically rotated overlapping parallel lines with enhanced reconstruction (PROPELLER). The signal intensity of flow were measured and regarded as the degree of FHVs. Simple and multiple linear regression analyses were applied to evaluate the association between the degree of FHVs and varying MR parameters as well as flow velocities.
Results:
On univariate analysis, PROPELLER technique (R(2 )= 0.448) demonstrated strongest correlation with the degree of FHV, followed by flow velocities (R(2 )= 0.204), FA (R(2 )= 0.126), and TE (R(2 )= 0.031), whereas TR/TI showed no significant correlations. On multivariate analysis, TE, FA, PROPELLER technique, and flow velocities were independent factors influencing the degree of FHVs (<0.001).
Conclusion:
Flow velocities, FA of refocusing pulse, TE, and PROPELLER technique significantly affected the degree of FHVs. Optimized MR parameters should be used consistently in future studies, which may provide more reliable results.