Comparison of a fluorometric assay kit with high-performance liquid chromatography for the assessment of serum retinol concentration
Related Experiment Video
Updated: Apr 8, 2026

Author Spotlight: Investigating Physiological Functions of Vitamin A Transporters Using HPLC-Based Vitamin A Profiling
Published on: December 27, 2024
Aglago Kouassivi Elom1, El Menchawy Imane1, Benjeddou Kaoutar1
1Joint Unit of Research in Nutrition and Food Sciences, URAC 39, Ibn Tofail University-CNESTEN.
Background:
Although high-performance liquid chromatography (HPLC) is the commonly used method for the analysis of retinol in biological samples, simple and rapid test kits are available.
Objectives:
This study compared a rapid test kit (ICHECK Fluoro®) to HPLC for the assessment of serum retinol concentrations.
Methods:
For the analysis by HPLC, sample preparation included standard deproteinization and extraction phases. The analysis by ICHECK was performed by injecting serum into IEX reagent vials (n=89) and mixing manually for separation. After precipitation of the proteins, the vial was introduced into the chamber of the ICHECK Fluoro and analysed at 0 min (ICHECK0min) and 15 min later (ICHECK15min). Bland and Altman approach was applied to test the agreement between HPLC and ICHECK.
Results:
Mean HPLC, ICHECK0min and ICHECK15min values were 421.2±106.0 µg/L, 423.1±118.3 µg/L and 413.2±107.6 µg/L, respectively. Retinol concentrations significantly decreased in the IEX solution over time (p<0.001). No significant proportional bias was observed between HPLC and ICHECK0min (r-0.038, p=0.73) and ICHECK15min (r=-0.024, p=0.82). Fixed biases (HPLC minus ICHECK) for ICHECK0min and ICHECK15min were respectively -1.9±23.1 µg/l (p=0.45) and 8.0±22.7 µg/l (p=0.002).
Conclusion:
ICHECK Fluoro may offer a reliable mean for assessing serum retinol for measurements performed with no significant time delay.