Related Experiment Video
Updated: Apr 8, 2026

Identification of Rare Bacterial Pathogens by 16S rRNA Gene Sequencing and MALDI-TOF MS
Published on: July 11, 2016
Direct Identification of Urinary Tract Pathogens From Urine Samples Using the Vitek MS System Based on
Yeongsic Kim1, Kang Gyun Park1, Kyungwon Lee2
1Department of Laboratory Medicine, College of Medicine, The Catholic University of Korea, Seoul, Korea.
Background:
We evaluated the coincidence rate between Vitek MS system (bioMérieux, France) and Vitek 2 in identifying uropathogens directly from urine specimens.
Methods:
Urine specimens submitted to our microbiology laboratory between July and September 2013 for Gram staining and bacterial culture were analyzed. Bacterial identification was performed by using the conventional method. Urine specimens showing a single morphotype by Gram staining were processed by culturing and matrix-assisted laser desorption ionization-time of flight mass spectrometry (MALDI-TOF MS). Of 2,370 urine specimens, 251 showed a single morphotype on Gram staining, and among them, 202 were available for MALDI-TOF MS.
Results:
In these 202 specimens, colony growth was observed in 189 specimens, and 145 specimens had significant growth of single-colony morphotype in culture. One hundred and ten (75.9%) of them had colony counts of ≥10(5) colony-forming units (CFU)/mL and included 71 enteric gram-negative bacteria (GNB), 5 glucose-non-fermenting GNB, 9 gram-positive cocci (GPC), and 25 yeasts. Furthermore, 70 (98.6%), 3 (60.0%), 4 (44.4%), and 5 (20.0%), respectively, of these were correctly identified by Vitek MS. Thirty-one specimens (21.4%; 11 GNB, 7 GPC, 12 yeasts, and 1 gram-positive bacillus) had colony counts of 10(4)-10(5) CFU/mL. Four specimens (2.8%) yielded colony counts of 10(3)-10(4) CFU/mL.
Conclusions:
Vitek MS showed high rate of accuracy for the identification of GNB in urine specimens (≥10(5) CFU/mL). This could become a rapid and accurate diagnostic method for urinary tract infection caused by GNB. However, for the identification of GPC and yeasts, further studies on appropriate pre-treatment are warranted.
More Related Videos
09:07Author Spotlight: Accelerating Diagnostic Accuracy with Direct Identification of Gram-Negatives from Blood Culture Bottles
Published on: May 24, 2024
11:25Preparation of a Blood Culture Pellet for Rapid Bacterial Identification and Antibiotic Susceptibility Testing
Published on: October 15, 2014
Related Concept Videos
Automated Microbial Diagnostics
MALDI-TOF Mass Spectrometry
Matrix-Assisted Laser Desorption Ionization (MALDI)
Microbiota of the Urogenital Tract
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...