Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
Sleep Transcutaneous vs. End-Tidal CO2 Monitoring for Patients with Neuromuscular Disease
Yu Hui Won1, Won Ah Choi, Jang Woo Lee
1From the Department of Physical Medicine and Rehabilitation, Research Institute of Clinical Medicine, and Institute for Medical Science, Chonbuk National University Medical School and Hospital, Jeonju, South Korea (YHW); Department of Rehabilitation Medicine, Gangnam Severance Hospital, Rehabilitation Institute of Neuromuscular Disease, Yonsei University College of Medicine, Seoul, South Korea (WAC, JWL, S-WK); Department of Physical Medicine and Rehabilitation, Rutgers New Jersey Medical School, University Hospital, Newark, New Jersey (JRB); Department of Physical Medicine and Rehabilitation, SRC Rehabilitation Hospital, Gwangju, South Korea (JP); and Yonsei University Graduate School of Medicine, Seoul, South Korea (YHW, S-WK).
Objective:
This study compared transcutaneous carbon dioxide partial pressure (PtcCO2) and end-tidal carbon dioxide partial pressure (PetCO2) monitoring during sleep for patients with neuromuscular disease.
Design:
This is a retrospective study of patients whose PtcCO2 and PetCO2 were monitored before they began using noninvasive mechanical ventilation. The outcomes were divided into four groupings: group 1, both PtcCO2 and PetCO2 are greater than or equal to 49 mm Hg; group 2, PtcCO2 is greater than or equal to 49 mm Hg but PetCO2 is less than 49 mm Hg; group 3, PtcCO2 is less than 49 mm Hg but PetCO2 is greater than or equal to 49 mm Hg; and group 4, both PtcCO2 and PetCO2 are less than 49 mm Hg.
Results:
A total of 39 subjects (mean [SD] age, 27.7 [19.3] yrs) were enrolled. PtcCO2 values were significantly higher than PetCO2 values (P < 0.001). The intraclass correlation coefficient between maximal and mean values of PtcCO2 and PetCO2 was 0.612 and 0.718, respectively. Bias and limits of agreement between PtcCO2 and PetCO2 were -7.5 mm Hg and -21.3 to 6.3 mm Hg for maximal values and -4.8 mm Hg and -14.8 to 5.3 mm Hg for mean values. Group 2 included 19 (48.7%) and group 3 included 3 (7.6%) patients who showed discrepancy of hypercapnia between two methods.
Conclusions:
Maximum PtcCO2 was significantly greater than maximum PetCO2 for both groups and, therefore, tends to be higher than PetCO2 in this population. This should be taken into consideration when assessing patients for sleep hypoventilation.