Impact of Number of Repeated Scans on Model Observer Performance for a Low-contrast Detection Task in CT
Chi Ma1, Lifeng Yu1, Baiyu Chen1
1Department of Radiology, Mayo Clinic, Rochester, MN.
Abstract:
Channelized Hotelling observer (CHO) has been validated against human observers for detection/classification tasks in clinical CT and shows encouraging correlations. However, the goodness of correlations depends on the number of repeated scans used in CHO to estimate the template and covariance matrices. The purpose of this study is to investigate how the number of repeated scans affects the CHO performance in predicting human observers. A phantom containing 21 low-contrast objects (3 contrast levels and 7 sizes) was scanned on a 128-slice CT scanner at three dose levels. Each scan was repeated 100 times. Images were reconstructed using a filtered-backprojection kernel and a commercial iterative reconstruction method. For each dose level and reconstruction setting, the low-contrast detectability, quantified with the area under receiver operating characteristic curve (Az), was calculated using a previously validated CHO. To determine the dependency of CHO performance on the number of repeated scans, the Az value was calculated for each object and dose/reconstruction setting using all 100 repeated scans. The Az values were also calculated using randomly selected subsets of the scans (from 10 to 90 scans with an increment of 10 scans). Using the Az from the 100 scans as the reference, the accuracy of Az from a smaller number of scans was determined. The minimum necessary number of scans was subsequently derived. For the studied signal-known-exactly detection task, results demonstrated that, the minimal number of scans required to accurately predict human observer performance depends on dose level, object size and contrast level, and channel filters.
More Related Videos
06:33Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
11:49Using High Resolution Computed Tomography to Visualize the Three Dimensional Structure and Function of Plant Vasculature
Published on: April 5, 2013
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...