Diffusion-Weighted Imaging of the Prostate: Image Quality and Geometric Distortion of Readout-Segmented Versus

Borna K Barth1, Alexander Cornelius, Daniel Nanz

  • 1From the *Institute of Diagnostic and Interventional Radiology, University Hospital Zürich, Zurich; †Department for Radiology, Cantonal Hospital Aarau, Aarau; and ‡Department of Urology, University Hospital Zürich, Zurich, Switzerland.

Abstract

Insights

Selective-excitation accelerated reduced-field of view (FOV) DWI (sTX-DWI) offers faster prostate imaging with similar quality but increased geometric distortion compared to readout-segmented DWI (rs-DWI). This trade-off is crucial for clinical application.

Area of Science:

  • Radiology and Imaging Science
  • Medical Physics
  • Oncology Imaging

Background:

  • Diffusion-weighted imaging (DWI) is vital for prostate cancer detection and characterization.
  • Traditional DWI techniques like readout-segmented DWI (rs-DWI) offer high quality but are time-consuming.
  • Accelerated DWI methods are needed to improve patient comfort and workflow efficiency.

Purpose of the Study:

  • To compare the image quality and geometric distortion of prostate rs-DWI and selective-excitation accelerated reduced-FOV DWI (sTX-DWI).
  • To evaluate the trade-offs between acquisition speed and image fidelity for these two DWI techniques.

Main Methods:

  • Sixty-five patients underwent 3-Tesla MRI including both rs-DWI and sTX-DWI sequences.
  • Image quality and geometric distortion were assessed by two readers using a 5-point Likert scale.
  • Quantitative geometric distortion was measured by comparing prostate diameters to T2-weighted imaging.

Main Results:

  • No significant differences were found in resolution, capsule demarcation, or zonal anatomy between rs-DWI and sTX-DWI.
  • Overall image quality was rated similarly high for both techniques by both readers.
  • sTX-DWI demonstrated significantly higher geometric distortion than rs-DWI, with more pronounced overestimation of prostate diameter.

Conclusions:

  • sTX-DWI allows for over 5-fold faster acquisition of prostate images compared to rs-DWI.
  • Image quality is comparable between sTX-DWI and rs-DWI.
  • The accelerated sTX-DWI technique comes at the cost of significantly increased geometric distortion.