Pulse Wave Velocity Testing in the Baltimore Longitudinal Study of Aging
Published on: February 7, 2014
Carotid-femoral pulse wave velocity assessment by two different methods: implications for risk assessment
Gernot Pichler1, Fernando Martinez, Antonio Vicente
1aHypertension Clinic. Department of Internal Medicine, Clinical Hospital of Valencia, INCLIVA, University of Valencia, Valencia bCIBEROBn, Carlos III Health Institute, Madrid, Spain.
Introduction:
Several devices are available for carotid-femoral pulse wave velocity (cfPWV) measurement, and a cut-off value for reference cfPWV has been established. However, discrepancies between devices have been reported.
Objectives:
The aim of the study was to establish the concordance of two common techniques (Complior and SphygmoCor), taking into account the anatomical distance between the measurement sites, and to investigate the impact on cardiovascular risk stratification.
Methods:
cfPWV, central and peripheral blood pressure were assessed in patients attending the hypertension outpatient clinic. The subtracted carotid-femoral distance was estimated both according to the manufacturer's recommendations and correcting the obtained values by 10.3%. Bland-Altman plots, Pearson's correlation coefficient, Lin's concordance correlation coefficient and multivariate models were used to investigate the difference in cfPWV.
Results:
cfPWV assessed in 118 patients (age 55 ± 12 years, 61% hypertensive patients, BMI 28.9 ± 4.4 kg/m2) with the Complior device was lower than that assessed with the SphygmoCor device, regardless of correcting the subtracted carotid-femoral distance (8.7 vs. 10.3 m/s and 9.3 m/s, respectively; P value < 0.001). The average difference was -1.59 ± 1.5 and -0.617 ± 1.39 m/s for corrected and uncorrected SphygmoCor values, respectively, SBP, BMI and female being the main determinants of the difference. Cardiovascular risk stratification changed in up to 40% of the study population, depending on the device and the arterial distance estimation.
Conclusion:
The concordance between the Complior and the SphygmoCor device is poor when the anatomical artery length is controlled for and in the presence of cardiovascular risk factors, resulting in a difference in classification of cardiovascular risk.
Related Concept Videos
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Pulse Assessment Sites
Pulse
The pulse serves as a clinical...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.