Imaging of Biological Tissues by Desorption Electrospray Ionization Mass Spectrometry
Published on: July 12, 2013
Electrospray droplet impact secondary ion mass spectrometry using a vacuum electrospray source
R Takaishi1, Y Sakai1, K Hiraoka1
1Clean Energy Research Center, University of Yamanashi, Takeda-4, Kofu, 400-8511, Japan.
Rationale:
In electrospray droplet impact (EDI) developed in our laboratory, an atmospheric pressure electrospray source has been used. To increase the ion beam intensity and reduce the evacuation load, a vacuum electrospray cluster ion source using a silica capillary was developed.
Methods:
A silica capillary with a tip inner diameter of 8 µm was used for vacuum electrospray using aqueous 10% methanol. To stabilize the flow rate of the liquid for nano-electrospray, a home-made constant pressure liquid pump was also developed.
Results:
By using the silica tip nano-electrospray emitter and a constant pressure pump, stable electrospray with flow rate of 22 nL/min was realized without using any heating system such as laser irradiation. Comparative study of mass spectra obtained by atmospheric pressure EDI (A-EDI) and vacuum EDI (V-EDI) was made for various samples such as thermometer molecule, peptide, polystyrene, Alq(3), NPD, C(60), indium, and SiO(2). V-EDI showed slightly milder ionization than A-EDI.
Conclusions:
Because V-EDI gave higher target current (5-10 nA) than A-EDI (a few nA at most), V-EDI secondary ion mass spectrometry (SIMS) would be a useful technique for the surface and interface analysis.
More Related Videos
Related Concept Videos
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
Chemical Ionization (CI) Mass Spectrometry
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
MALDI-TOF Mass Spectrometry
Mass Spectrum: Interpretation
Mass Spectrometers