The SOMNOtouch device as a novel method for measuring short-term blood pressure variability: a comparison with the
Lisa S Manning1, Thompson G Robinson, Ronney B Panerai
1University of Leicester, Department of Cardiovascular Sciences and NIHR Biomedical Research Unit in Cardiovascular Disease, Leicester, UK.
Background:
Current noninvasive techniques to capture short-term blood pressure variability (BPV) have methodological and practical limitations. This study assessed the ability of a novel device, the SOMNOtouch, which derives continuous blood pressure (BP) measures from pulse transit time, to estimate BPV, compared with the widely used Finometer.
Methods:
BP monitoring was performed simultaneously on the SOMNOtouch and Finometer devices in 16 healthy volunteers. Systolic and diastolic BPVs, defined as SD and coefficient of variation, were derived from measurements from each device for three predefined periods: 0-3, 7-10, and 0-10 min.
Results:
Agreement in BPV indices from the two devices was assessed using the Bland-Altman technique. For all BPV parameters, over all measurement periods, broad scatter was observed on Bland-Altman plots. Bias (limits of agreement) for minutes 0-10: SD of systolic BP, -3.03 mmHg (-10.88 to +4.55), SD of diastolic BP -1.65 mmHg (-4.41 to +1.11).
Conclusions:
The poor agreement observed in BPV estimates between the devices may reflect the inability of the current pulse transit time method to sensitively detect changes in BP. Further investigation is needed before such methods can be reliably used to measure short-term BPV.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Measurement of Blood Pressure
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.