Quantitative Fundus Autofluorescence for the Evaluation of Retinal Diseases
Published on: March 11, 2016
Quality evaluation of digital fundus images through combined measures
Diana Veiga1, Carla Pereira2, Manuel Ferreira1
1University of Minho , Centro Algoritmi, Azurém, Guimarães 4800-025, Portugal ; ENERMETER , Parque Industrial de Celeirós 2ª Fase, Lugar de Gaião, Lotes 5/6, Braga 4705-025, Portugal.
Abstract:
The evaluation of image quality is an important step before an automatic analysis of retinal images. Several conditions can impair the acquisition of a good image, and minimum image quality requirements should be present to ensure that an automatic or semiautomatic system provides an accurate diagnosis. A method to classify fundus images as low or good quality is presented. The method starts with the detection of regions of uneven illumination and evaluates if the segmented noise masks affect a clinically relevant area (around the macula). Afterwards, focus is evaluated through a fuzzy classifier. An input vector is created extracting three focus features. The system was validated in a large dataset (1454 fundus images), obtained from an online database and an eye clinic and compared with the ratings of three observers. The system performance was close to optimal with an area under the receiver operating characteristic curve of 0.9943.