Autocorrelation Analysis Combined with a Wavelet Transform Method to Detect and Remove Cosmic Rays in a Single Raman Spectrum
Related Experiment Video
Updated: Apr 7, 2026

Resolving Water, Proteins, and Lipids from In Vivo Confocal Raman Spectra of Stratum Corneum through a Chemometric Approach
Published on: September 26, 2019
Augusto Maury1, Reynier I Revilla
1Center for Advanced Studies of Cuba, Havana 19370, Cuba.
Abstract:
Cosmic rays (CRs) occasionally affect charge-coupled device (CCD) detectors, introducing large spikes with very narrow bandwidth in the spectrum. These CR features can distort the chemical information expressed by the spectra. Consequently, we propose here an algorithm to identify and remove significant spikes in a single Raman spectrum. An autocorrelation analysis is first carried out to accentuate the CRs feature as outliers. Subsequently, with an adequate selection of the threshold, a discrete wavelet transform filter is used to identify CR spikes. Identified data points are then replaced by interpolated values using the weighted-average interpolation technique. This approach only modifies the data in a close vicinity of the CRs. Additionally, robust wavelet transform parameters are proposed (a desirable property for automation) after optimizing them with the application of the method in a great number of spectra. However, this algorithm, as well as all the single-spectrum analysis procedures, is limited to the cases in which CRs have much narrower bandwidth than the Raman bands. This might not be the case when low-resolution Raman instruments are used.
Related Concept Videos
Raman Spectroscopy: Overview
However, a small fraction of the scattered light exhibits a frequency shift due to the exchange of energy between the incident photons and...
Raman Spectroscopy Instrumentation: Overview
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...
Attenuated Total Reflectance (ATR) Infrared Spectroscopy: Overview
The ATR process begins by directing a beam...
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
2D NMR: Homonuclear Correlation Spectroscopy (COSY)