Automated tube voltage adaptation in combination with advanced modeled iterative reconstruction in thoracoabdominal
Jan-Erik Scholtz1, Julian L Wichmann1, Kristina Hüsers1
1Department of Diagnostic and Interventional Radiology, University Hospital Frankfurt, Theodor-Stern-Kai 7, 60590 Frankfurt, Germany.
Rationale And Objectives:
To evaluate image quality and radiation exposure of portal venous-phase thoracoabdominal third-generation 192-slice dual-source computed tomography (DSCT) with automated tube voltage adaptation (TVA) in combination with advanced modeled iterative reconstruction (ADMIRE).
Materials And Methods:
Fifty-one patients underwent oncologic portal venous-phase thoracoabdominal follow-up CT twice within 7 months. The initial examination was performed on second-generation 128-slice DSCT with fixed tube voltage of 120 kV in combination with filtered back projection reconstruction. The second examination was performed on a third-generation 192-slice DSCT using automated TVA in combination with ADMIRE. Attenuation and image noise of liver, spleen, renal cortex, aorta, vena cava inferior, portal vein, psoas muscle, and perinephric fat were measured. Signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were calculated. Radiation dose was assessed as size-specific dose estimates (SSDE). Subjective image quality was assessed by two observers using five-point Likert scales. Interobserver agreement was calculated using intraclass correlation coefficients (ICC).
Results:
Automated TVA set tube voltage to 90 kV (n = 8), 100 kV (n = 31), 110 kV (n = 11), or 120 kV (n = 1). Average SSDE was decreased by 34.9% using 192-slice DSCT compared to 128-slice 120-kV DSCT (7.8 ± 2.4 vs. 12.1 ± 3.2 mGy; P < .001). Image noise was substantially lower; SNR and CNR were significantly increased in 192-slice DSCT compared to 128-slice DSCT (all P < .005). Image quality was voted excellent for both acquisition techniques (5.00 vs. 4.93; P = .083).
Conclusions:
Automated TVA in combination with ADMIRE on third-generation 192-slice DSCT in portal venous-phase thoracoabdominal CT provides excellent image quality with reduced image noise and increased SNR and CNR, whereas average radiation dose is reduced by 34.9% compared to 128-slice DSCT.
More Related Videos
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
09:57Development and Evaluation of 3D-Printed Cardiovascular Phantoms for Interventional Planning and Training
Published on: January 18, 2021
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...