[The Mitroflow bioprosthesis. Results after 6 years]
Revista Espanola De Cardiologia
|November 1, 1989
Summary
The Mitroflow pericardial heart valve shows satisfactory medium-term performance, with good survival rates. However, early primary tissue failures were noted in the mitral position, requiring further investigation.
Area of Science:
- Cardiovascular Surgery
- Biomaterials Science
- Medical Device Engineering
Context:
- The study evaluates the clinical performance of the Mitroflow pericardial heart valve.
- 139 valves were implanted in 129 patients between March 1982 and December 1986.
- Patient demographics included a mean age of 55.2 years, with 69 males and 60 females.
Purpose:
- To assess the medium-term clinical performance and durability of the Mitroflow pericardial heart valve.
- To report on outcomes including mortality, thromboembolism, tissue failure, and reoperation rates.
- To identify any specific complications associated with the valve's use.
Summary:
- Overall hospital mortality was 1.6%, with 3 late deaths (2.4%).
- The linearized incidence of late mortality was 0.49% per patient/year, with 96% +/- 1.8 actuarial survival.
- Thromboembolism incidence was 1.22% per patient/year (93% +/- 2.7 freedom rate).
- Three patients experienced primary tissue failure in the mitral position (0.73% linearized incidence, 95% +/- 3.2 freedom rate).
- Four patients required reoperation (0.49% per patient/year).
Impact:
- The Mitroflow pericardial heart valve demonstrates satisfactory medium-term clinical performance.
- A significant rate of early primary tissue failure was observed specifically in the mitral position.
- Findings highlight the need for continued monitoring and potential design improvements for mitral valve applications.