Accuracy Evaluation of CONTOUR(®)PLUS Compared With Four Blood Glucose Monitoring Systems
Nancy Dunne1, Maria T Viggiani2, Scott Pardo2
1Bayer HealthCare LLC, Diabetes Care, Whippany, NJ, USA. nancy.dunne@bayer.com.
Introduction:
The aim of this study was to compare the accuracy of 5 blood glucose monitoring systems (BGMSs; CONTOUR(®)PLUS [CP], Accu-Chek(®) Active [ACA], Accu-Chek(®) Performa [ACP], FreeStyle Freedom™ [FF], OneTouch(®) SelectSimple™ [OTSS]).
Methods:
Study staff tested fingerstick samples from 106 subjects aged ≥18 years using the 5 BGMSs. Some samples were modified to achieve blood glucose concentrations throughout the measuring range. The primary endpoint was comparison of the mean absolute relative difference (MARD) from the reference value (Yellow Springs Instruments [YSI]) across the overall tested glucose range. Other endpoints were MARD in the low (≤80 mg/dL [≤4.4 mmol/L]), middle (81-180 mg/dL [4.5-10.0 mmol/L]), and high (>180 mg/dL [>10.0 mmol/L]) glucose ranges, and MARD for unmodified samples in the overall glucose range.
Results:
CONTOUR(®)PLUS had a statistically significantly lower MARD than all BGMSs across the overall tested range (27-460 mg/dL [1.5-25.5 mmol/L]) and in the high glucose range. In the low glucose range, CP had a lower MARD than all BGMSs, which was statistically significant except for ACP. For unmodified samples across the overall tested range, CP had a lower MARD than all BGMSs and was statistically significantly lower except for ACA.
Conclusions:
CONTOUR(®)PLUS had the lowest mean difference from the reference values (by MARD) when compared with other BGMSs across multiple glucose ranges with modified and unmodified samples.
Funding:
Bayer HealthCare LLC, Diabetes Care.
Trial Registration Number:
ClinicalTrials.gov Identifier NCT01714232.
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Assessment of blood pressure in brachial artery(two-step method)
Measurement of Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: