Related Experiment Video
Updated: Apr 7, 2026

10:34
FRET Microscopy for Real-time Monitoring of Signaling Events in Live Cells Using Unimolecular Biosensors
Published on: August 20, 2012
23.9K
A Practical Solution for 77 K Fluorescence Measurements Based on LED Excitation and CCD Array Detector
Jacob Lamb1, Kristin Forfang1, Martin Hohmann-Marriott1
1Department of Biotechnology & PhotoSynLab, Norwegian University of Science and Technology, Trondheim, Norway.
Plos One
|July 16, 2015
Summary
Researchers developed a portable, low-cost instrument for measuring 77K fluorescence spectra in photosynthetic microorganisms. This new method provides crucial insights into photosynthetic machinery organization, previously requiring expensive equipment.
Area of Science:
- Photosynthesis research
- Biophysical techniques
- Spectroscopy
Background:
- Fluorescence emission spectra at 77 K reveal details of photosynthetic machinery organization in bacteria and eukaryotes.
- Conventional methods require large, expensive table-top fluorometers, limiting accessibility.
- Recent advancements offer portable, reliable, and low-maintenance spectroscopic instruments.
Purpose of the Study:
- To demonstrate the construction of a cost-effective, portable instrument for recording 77 K fluorescence spectra.
- To enable detailed analysis of photosynthetic machinery organization in microorganisms.
Main Methods:
- Assembly of a custom instrument using a low-power monochromatic light-emitting diode (LED) and a portable CCD array spectrometer.
- Integration of optical components via fiber optic cable and a custom housing supporting a dewar flask.
- Utilized the instrument to record chlorophyll fluorescence emission spectra.
Main Results:
- Successfully built and demonstrated a portable, low-cost instrument for 77 K fluorescence spectroscopy.
- Reliably determined chlorophyll fluorescence emission spectra for Synechocystis sp. PCC 6803 (cyanobacterium) and Chlamydomonas reinhardtii (green alga).
- The developed instrument provides an accessible alternative to conventional, costly fluorometers.
Conclusions:
- The developed portable spectroscopic instrument is effective for analyzing 77 K fluorescence emission spectra of photosynthetic microorganisms.
- This approach democratizes the study of photosynthetic machinery organization, making advanced spectroscopy more accessible.
- The instrument facilitates reliable spectral determination for key model organisms in photosynthesis research.