Ionic conductivity of peritoneal dialysate: a new, easy and fast method of assessing peritoneal membrane function in
Vincenzo La Milia1, Giuseppe Pontoriero1, Giovambattista Virga2
1Division of Nephrology and Dialysis, A. Manzoni Hospital, Lecco, Italy.
Background:
Peritoneal membrane function can be assessed using the peritoneal equilibration test (PET) and similar tests, but these are almost always complicated to use, require a considerable amount of working time and their results cannot always be easily interpreted. Ionic conductivity is a measure of the ability of an electrolyte solution to conduct electricity. We tested the hypothesis that the ionic conductivity of peritoneal dialysate can be used to evaluate peritoneal membrane function in peritoneal dialysis patients.
Methods:
We measured the ionic conductivity and classic biochemical parameters of peritoneal dialysate in 69 patients during a modified PET and compared their ability to evaluate peritoneal membrane function and to diagnose ultrafiltration failure (UFF).
Results:
Ionic conductivity was correlated well with classical parameters of peritoneal transport as glucose reabsorption of glucose (D/D0: r(2) = 0.62, P < 0.0001) and creatinine transport (D/PCreat: r(2) = 0.72, P < 0.0001). Twelve patients (17%) experienced UFF and, in them, the ionic conductivity area under the receiver-operating characteristic curve was 0.91 (95% confidence interval: 0.81-0.96) with sensitivity of 1.00 and specificity of 0.84 at a cut-off value of 12.75 mS/cm.
Conclusions:
These findings indicate that the ionic conductivity of peritoneal dialysate can be used as a new screening tool to evaluate peritoneal membrane function.
More Related Videos
13:30Use of the Ramsay Assay to Measure Fluid Secretion and Ion Flux Rates in the Drosophila melanogaster Malpighian Tubule
Published on: November 25, 2015
08:30Preparation and Utilization of Freshly Isolated Human Detrusor Smooth Muscle Cells for Characterization of 9-Phenanthrol-Sensitive Cation Currents
Published on: January 31, 2020
Related Concept Videos
Peritoneal Dialysis I: Introduction and Procedure
Dialysis
Potentiometry: Membrane Electrodes
Electrical Transport
Debye–Huckel–Onsager Conductance Equation
Peritoneal Dialysis III: Nursing Management