Assessment of implant position after total knee arthroplasty by dual-energy computed tomography
Ferdinando Ferrara1, Antonio Cipriani2, Santi Rapisarda2
1Department of Radiological Sciences, Catholic University of the Sacred Heart, "A. Gemelli" Hospital, Rome, Italy ferdinandoferrara1@gmail.com.
Background:
Correct alignment of prosthetic components is the most important factor for the success of total knee arthroplasty (TKA). Dual-energy computed tomography (DECT) may be a reliable method in determining implant position after TKA.
Purpose:
To evaluate the accuracy and reproducibility of DECT in determining implant position after TKA.
Material And Methods:
Institutional review board approval was obtained. Forty-five patients (age 75.2 ± 6.4 years) prospectively underwent TKA at our institution between May and December 2012. DECT was performed 1 year after surgery, using an alignment similar to a standing position and generating images at an extrapolated energy of 120 kVp, in order to reduce metal artifacts. Implant position was evaluated by two independent readers. Intra- and inter-observer agreements were calculated. DECT measurements on implant position were compared with the preoperative planning based on radiographs. Additional clinical and DECT findings were taken into account.
Results:
Metal artifact reduction was judged satisfactory in all cases. Regarding implant position assessed with DECT, good to excellent intra-observer (k: 0.74-0.87 and k: 0.75-0.88, respectively), and inter-observer agreement (k: 0.72-0.82) were found. In the comparison with preoperative planning, the widest limits of agreement were within 3.9° for the sagittal orientation of tibial component. A single patient with postoperative knee pain and stiffness had periprosthetic osteopenia, quadriceps femoris tendon calcifications, articular effusion, and excessive intrarotation of the femoral component.
Conclusion:
DECT is an accurate and reproducible tool for determining implant position after TKA.