On the Design and Implementation of a Highly Accurate Pulse Predictor for Exercise Equipment
Goal:
This study aims to develop highly accurate heart rate monitoring from the hand-held contact signal within a noisy environment during exercise.
Methods:
The periodic pattern and uncertainties of a physiological signal are modeled by a Laplacian random process. Based on this statistical model, a highly accurate pulse predictor (HAPPEE) is derived and implemented in real-time on a Cypress PSoC 5LP development board. A real-time experiment is designed to compare HAPPEE with a commercial heart rate monitor from POLAR. The percentage of credible estimates and the mean square error (MSE) of credible estimates are reported for experiments with seven healthy subjects.
Results:
The overall percentage of credible estimates is 99.2% for HAPPEE and 93.6% for POLAR. The overall MSE of credible estimates is 3.1 for HAPPEE and 7.7 for POLAR. These results show that HAPPEE is more accurate than POLAR.
Conclusion:
HAPPEE is able to accurately monitor heart rate within a noisy environment during exercise.
Significance:
Unlike existing heart rate estimation methods, HAPPEE does not require pulse detection or tuning parameters. It can be easily implemented in real-time on a low power and low cost development board for exercise equipment and outperforms a commercial heart rate monitor.
Related Concept Videos
Special considerations while measuring pulse
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Pulse
The pulse serves as a clinical...
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps:
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...