Comparison Among Methods of Retinopathy Assessment (CAMRA) Study: Smartphone, Nonmydriatic, and Mydriatic Photography
Martha E Ryan1, Ramachandran Rajalakshmi2, Vijayaraghavan Prathiba2
1Department of Ophthalmology, Emory Eye Center, Emory University School of Medicine, Atlanta, Georgia.
Purpose:
We compared smartphone fundus photography, nonmydriatic fundus photography, and 7-field mydriatic fundus photography for their abilities to detect and grade diabetic retinopathy (DR).
Design:
This was a prospective, comparative study of 3 photography modalities.
Participants:
Diabetic patients (n = 300) were recruited at the ophthalmology clinic of a tertiary diabetes care center in Chennai, India.
Methods:
Patients underwent photography by all 3 modalities, and photographs were evaluated by 2 retina specialists.
Main Outcome Measures:
The sensitivity and specificity in the detection of DR for both smartphone and nonmydriatic photography were determined by comparison with the standard method, 7-field mydriatic fundus photography.
Results:
The sensitivity and specificity of smartphone fundus photography, compared with 7-field mydriatic fundus photography, for the detection of any DR were 50% (95% confidence interval [CI], 43-56) and 94% (95% CI, 92-97), respectively, and of nonmydriatic fundus photography were 81% (95% CI, 75-86) and 94% (95% CI, 92-96%), respectively. The sensitivity and specificity of smartphone fundus photography for the detection of vision-threatening DR were 59% (95% CI, 46-72) and 100% (95% CI, 99-100), respectively, and of nonmydriatic fundus photography were 54% (95% CI, 40-67) and 99% (95% CI, 98-100), respectively.
Conclusions:
Smartphone and nonmydriatic fundus photography are each able to detect DR and sight-threatening disease. However, the nonmydriatic camera is more sensitive at detecting DR than the smartphone. At this time, the benefits of the smartphone (connectivity, portability, and reduced cost) are not offset by the lack of sufficient sensitivity for detection of DR in most clinical circumstances.
Related Concept Videos
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Renal Drug Clearance: Comparison Between Renal Excretion Methods
Renal clearance is often associated with the renal glomerular filtration rate (GFR), which represents the rate at which plasma is filtered through the glomeruli in the kidney. When drug reabsorption is minimal and there is no active secretion, renal clearance is closely related to the...
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Evolutionary Relationships through Genome Comparisons
Comparison Between Electrical And Gravitational Forces
Since both are inverse square law forces, the distance gets canceled when the ratio of the two forces is considered. Instead, the ratio of the electrical and gravitational forces depends on...
Comparison between RL and RC circuits