Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
Computed Tomography Number Measurement Consistency Under Different Beam Hardening Conditions: Comparison Between
Tian He1, Xiaojun Qian, Renyou Zhai
1From the Department of Radiology, Beijing Chaoyang Hospital, Affiliated to Capital Medical University, Beijing, China.
Purpose:
To compare computed tomography (CT) number measurement consistency under different beam hardening conditions in phantom experiment between dual-energy spectral CT and conventional CT imaging.
Materials And Methods:
A phantom with 8 cells in periphery region and 1 cell in central region were used. The 8 conditioning tubes in the periphery region were filled with 1 of the 3 iodine solutions to simulate different beam hardening conditions: 0 for no beam hardening (NBH), 20 mg/mL for weak beam hardening (WBH) and 50 mg/mL for severe beam hardening (SBH) condition. Test tube filled with 0, 0.1, 0.5, 1, 2, 5, 10, 20, and 50 mg/mL iodine solution was placed in the central cell alternately. The phantom was scanned with conventional CT mode with 80, 100, 120, and 140 kVp and dual energy spectral CT mode. For spectral CT, 11 monochromatic image sets from 40 to 140 keV with interval of 10 keV were reconstructed. The CT number shift caused by beam hardening was evaluated by measuring the CT number difference (ΔCT) with and without beam hardening, with the following formulas: ΔCTWBH = |CTWBH - CTNBH| and ΔCTSBH = |CTSBH - CTNBH|. Data were compared with 1-way analysis of variance.
Results:
Under both WBH and SBH conditions, the CT number shifts in all monochromatic image sets were less than those for polychromatic images (all P < 0.001). Under WBH condition, the maximum CT number shift was less than 6 Hounsfield units for monochromatic spectral CT images of all energy levels; under SBH condition, only monochromatic images at 70 keV and 80 keV had CT number shift less than 6 HU.
Conclusion:
Dual energy spectral CT imaging provided more accurate CT number measurement than conventional CT under various beam hardening conditions. The optimal keV level for monochromatic spectral CT images with the most accurate CT number measurement depends on the severities of beam hardening condition.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...