Monitoring Dead Space in Mechanically Ventilated Children: Volumetric Capnography Versus Time-Based Capnography
Anoopindar K Bhalla1, Sarah Rubin2, Christopher J L Newth2
1Department of Anesthesiology and Critical Care Medicine, Children's Hospital Los Angeles, Los Angeles, California. Keck School of Medicine, University of Southern California, Los Angeles, California. abhalla@chla.usc.edu.
Background:
Volumetric capnography dead-space measurements (physiologic dead-space-to-tidal-volume ratio [VD/VT] and alveolar VD/VT) are considered more accurate than the more readily available time-based capnography dead-space measurement (end-tidal alveolar dead-space fraction [AVDSF]). We sought to investigate the correlation between volumetric capnography and time-based capnography dead-space measurements.
Methods:
This was a single-center prospective cohort study of 65 mechanically ventilated children with arterial lines. Physiologic VD/VT, alveolar VD/VT, and AVDSF were calculated with each arterial blood gas using capnography data.
Results:
We analyzed 534 arterial blood gases from 65 children (median age 4.9 y, interquartile range 1.7-12.8). The correlation between physiologic VD/VT and AVDSF (r = 0.66, 95% CI 0.59-0.72) was weaker than the correlation between alveolar VD/VT and AVDSF (r = 0.8, 95% CI 0.76-0.85). The correlation between physiologic VD/VT and AVDSF was weaker in children with low PaO2 /FIO2 (< 200 mm Hg), low exhaled VT (< 100 mL), a pulmonary reason for mechanical ventilation, or large airway VD (> 3 mL/kg). All 3 dead-space measurements were highly correlated (r > 0.7) in children without hypoxemia (PaO2 /FIO2 > 300 mm Hg), mechanically ventilated for a neurologic or cardiac reason, or on significant inotropes or vasopressors.
Conclusions:
In mechanically ventilated children without significant hypoxemia or with cardiac output-related dead-space changes, physiologic VD/VT was highly correlated with AVDSF and alveolar VD/VT. In children with significant hypoxemia, physiologic VD/VT was poorly correlated with AVDSF. Alveolar VD/VT and AVDSF correlated well in most tested circumstances. Therefore, AVDSF may be useful in most children for alveolar dead-space monitoring.
More Related Videos
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Respiratory Volumes and Capacities
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Respiratory Volumes
Tidal Volume (TV) Tidal volume (TV) is the air inhaled or exhaled in a...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...