Adjustment of Open-Loop Settings to Improve Closed-Loop Results in Type 1 Diabetes: A Multicenter Randomized Trial
Eyal Dassau1, Sue A Brown1, Ananda Basu1
1Department of Chemical Engineering (E.D., R.G., J.B.L., H.C.Z., F.J.D.), University of California Santa Barbara, Santa Barbara, CA 93106; William Sansum Diabetes Center (E.D., J.E.P., R.G., J.B.L., K.C., W.C.B., P.K.B., H.C.Z., F.J.D.), Santa Barbara, CA 93105; Center for Diabetes Technology (S.A.B., S.P., D.L., M.M.-M., C.A.W., B.K.), University of Virginia, Charlottesville, VA 22904; Endocrine Research Unit (A.B., Y.C.K., L.H., A.M., V.D., S.K.M.-S.), Mayo Clinic, Rochester, MN 55905; and Department of Information Engineering (M.S., D.M., C.C.), University of Padova, 35131 Padua, Italy.
Context:
Closed-loop control (CLC) relies on an individual's open-loop insulin pump settings to initialize the system. Optimizing open-loop settings before using CLC usually requires significant time and effort.
Objective:
The objective was to investigate the effects of a one-time algorithmic adjustment of basal rate and insulin to carbohydrate ratio open-loop settings on the performance of CLC.
Design:
This study reports a multicenter, outpatient, randomized, crossover clinical trial.
Patients:
Thirty-seven adults with type 1 diabetes were enrolled at three clinical sites.
Interventions:
Each subject's insulin pump settings were subject to a one-time algorithmic adjustment based on 1 week of open-loop (i.e., home care) data collection. Subjects then underwent two 27-hour periods of CLC in random order with either unchanged (control) or algorithmic adjusted basal rate and carbohydrate ratio settings (adjusted) used to initialize the zone-model predictive control artificial pancreas controller. Subject's followed their usual meal-plan and had an unannounced exercise session.
Main Outcomes And Measures:
Time in the glucose range was 80-140 mg/dL, compared between both arms.
Results:
Thirty-two subjects completed the protocol. Median time in CLC was 25.3 hours. The median time in the 80-140 mg/dl range was similar in both groups (39.7% control, 44.2% adjusted). Subjects in both arms of CLC showed minimal time spent less than 70 mg/dl (median 1.34% and 1.37%, respectively). There were no significant differences more than 140 mg/dL.
Conclusions:
A one-time algorithmic adjustment of open-loop settings did not alter glucose control in a relatively short duration outpatient closed-loop study. The CLC system proved very robust and adaptable, with minimal (<2%) time spent in the hypoglycemic range in either arm.
Related Concept Videos
Open and closed-loop control systems
An open-loop control system operates without feedback from the output. It consists of two primary elements: the controller and the controlled process. The controller receives an input signal...
Feedback Loops
Diabetes Mellitus: Overview and Type I Subtype
Type 1 diabetes is an autoimmune disease in which the immune system mistakenly attacks and destroys the insulin-producing beta cells in the pancreas. As a result, the body is unable to produce sufficient insulin, and individuals with...
Diabetes: Management and Pharmacotherapy
Insulin remains the cornerstone of treatment for most patients with type 1 and many...
Insulin: Dosing Regimen and Adverse Effects
The basal dose constitutes about 40%-50% of the total daily dose, with the rest as premeal insulin. The mealtime insulin dose should mirror...
Carbohydrate Metabolism
Starch accounts for approximately 60% of the carbohydrates consumed by humans. Since amylase enzymes cannot function in the stomach's acidic environment, starch can only be digested in the mouth and small intestine. Simple sugars are found naturally in milk and fruits in...