Related Experiment Video
Updated: Apr 6, 2026

Author Spotlight: Assessment of Cardiac Output Calculation by Thermodilution in Pigs for Effective Perfusion Flow During EVLP
Published on: June 28, 2024
Ability of esCCO to track changes in cardiac output
M Biais1, R Berthezène2, L Petit2
1Department of Anaesthesiology and Critical Care III, Bordeaux University Hospital, F-33000 Bordeaux, France INSERM, Adaptation Cardiovasculaire à L'ischémie, U1034, F-33600 Pessac, France Univ. Bordeaux, Adaptation Cardiovasculaire à L'ischémie, U1034, F-33600 Pessac, France matthieu.biais@chu-bordeaux.fr.
Background:
We investigated whether cardiac output measured with pulse wave transit time (esCCO, Nihon Kohden, Tokyo, Japan) is able to track changes in cardiac output induced by an increase in preload (volume expansion/passive leg-raising) or by changes in vasomotor tone (variation in norepinephrine dosage) in critically ill patients.
Methods:
Eighty patients for whom the decision to give fluid (500 mL of saline over 15 min) (n=20), to perform passive leg-raising (n=20), and to increase (n=20) or to decrease (n=20) norepinephrine were included by the physician. Cardiac output was measured with pulse wave transit time (CO-esCCO) and transthoracic echocardiography (CO-TTE) before and after therapeutic intervention.
Results:
Comparison between CO-TTE and CO-esCCO showed a bias of -0.7 l min(-1) and limits of agreement of -4.4 to 2.9 l min(-1), before therapeutic intervention and a bias of -0.5 l min(-1) and limits of agreement of -4.2 to 3.2 l min(-1) after therapeutic intervention. Bias was correlated with systemic vascular resistance (r(2)=0.60, P<0.0001). Percentage error was 61% before and 59% after therapeutic intervention. Considering the overall data (n=80), the concordance rate was 84%, polar plot analysis revealed an angular bias (sd) of -11°(35°) and radial limits of agreement of (sd 50°). With regard to passive leg-raising and volume expansion groups (n=40), the concordance rate was 83%, the angular bias (sd) was -20°(36°) and radial limits of agreement ( 50°). Considering variations in norepinephrine dosage groups (n=40), the concordance rate was 86%, the angular bias (sd) was -1.8°(33°) and radial limits of agreement (40°).
Conclusions:
esCCO was not able to track changes in cardiac output, induced by an increase in preload or by variations in vasomotor tone. Therefore, esCCO cannot guide haemodynamic interventions in critically ill patients.
Related Concept Videos
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac...
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Imbalances in Cardiac Output
CHF can occur due to the failure of either side of the heart. Left-side failure leads to pulmonary congestion—the right side continues to send...
Exercise and Cardiac Output
Sustained exercise increases the muscles' oxygen demand, which can be...