Patient Radiation Dose Reduction during Transarterial Chemoembolization Using a Novel X-Ray Imaging Platform
Ryan Kohlbrenner1, K Pallav Kolli1, Andrew G Taylor1
1Department of Radiology, University of California, San Francisco, 505 Parnassus Ave., M-391, San Francisco, CA 94143.
Purpose:
To evaluate radiation dose reduction in patients undergoing transarterial chemoembolization with the use of a new image acquisition and processing platform.
Materials And Methods:
Radiation-dose data were obtained from 176 consecutive chemoembolization procedures in 135 patients performed in a single angiography suite. From January 2013 through October 2013, 85 procedures were performed by using our institution's standard fluoroscopic settings. After upgrading the x-ray fluoroscopy system with an image acquisition and processing platform designed to reduce image noise and reduce skin entrance dose, 91 chemoembolization procedures were performed from November 2013 through December 2014. Cumulative dose-area product (CDAP), cumulative air kerma (CAK), and total fluoroscopy time were recorded for each procedure. Image quality was assessed by three interventional radiologists blinded to the x-ray acquisition platform used.
Results:
Patient radiation dose indicators were significantly lower for chemoembolization procedures performed with the novel imaging platform. Mean CDAP decreased from 3,033.2 dGy·cm(2) (range, 600.3-9,404.1 dGy·cm(2)) to 1,640.1 dGy·cm(2) (range, 278.6-6,779.9 dGy·cm(2); 45.9% reduction; P < .00001). Mean CAK decreased from 1,445.4 mGy (range, 303.6-5,233.7 mGy) to 971.7 mGy (range, 144.2-3,512.0 mGy; 32.8% reduction; P < .0001). A 20.3% increase in mean total fluoroscopy time was noted after upgrading the imaging platform, but blinded analysis of the image quality revealed no significant degradation.
Conclusions:
Although a small increase in fluoroscopy time was observed, a significant reduction in patient radiation dose was achieved by using the optimized imaging platform, without image quality degradation.
More Related Videos
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
09:49Dual-phase Cone-beam Computed Tomography to See, Reach, and Treat Hepatocellular Carcinoma during Drug-eluting Beads Transarterial Chemo-embolization
Published on: December 2, 2013
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System V: CT