Related Experiment Video
Updated: Apr 6, 2026

Translational Rabbit Model of Chronic Cardiac Pacing
Published on: January 6, 2023
Predictive factors of lead failure in patients implanted with cardiac devices
Yoshiyasu Aizawa1, Masachika Negishi1, Shin Kashimura1
1Department of Cardiology, Keio University School of Medicine, Tokyo, Japan.
Introduction:
Lead failures (LFs) are one of the most common complications in patients implanted with cardiovascular implantable electronic devices. LFs often cause serious secondary complications such as inappropriate ICD shocks or asystole. This study aimed to identify the clinical factors associated with the occurrence of LFs.
Methods:
A total of 735 consecutive device implantations (mean age 67±15years, males 64%) performed at a single university hospital setting from 1997 to 2014 were included. The implanted devices consisted of 421 pacemakers, 250 implantable cardioverter defibrillators (ICD), 9 cardiac resynchronization therapy pacemakers (CRT-P), and 55 CRT defibrillators (CRT-D). The primary endpoint was the development of an LF.
Results:
During a mean duration of 5.8±4.3years, 38 LFs developed in 31 patients (mean age 56±14years). LFs included 32 ICD (7 Sprint Fidelis, 2 Riata), and 6 pacing leads. Nine patients received inappropriate ICD shocks and 1 had syncope due to an LF. All patients underwent lead reinsertions with device replacements. Eight patients required opposite site implantations due to venous occlusions. The predictive factors of LFs were the age, male sex, taller body length, ICD vs. pacemaker, lesser lead number, extra-thoracic puncture of the axillary vein vs. a cut-down of the cephalic vein, use of recalled leads and patients with idiopathic ventricular fibrillation (IVF) and Brugada syndrome (BrS).
Conclusion:
LFs occurred mainly with ICD leads. A lesser age, the puncture method, lead model, and diagnosis of IVF/BrS were associated with the development of LFs.
More Related Videos
16:40A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
05:16Cutoff Value of Phase Angle by Bioelectrical Impedance Analysis at Admission as a Prognostic Factor in Patients with Acute Heart Failure
Published on: June 10, 2025
Related Concept Videos
Heart Failure I: Introduction
Pathophysiology of Heart Failure
Drug Toxicity: Risk factors