Quantification of remifentanil and propofol in human plasma: a LC-MS/MS assay validated according to the EMA
Jean Claude Alvarez1,2, Emuri Abe2, Isabelle Etting2
1Plateforme de spectrométrie de masse MasSpecLab, UFR des Sciences de la Santé Simone Veil, Université Versailles Saint Quentin, 2 avenue de la source de la Bièvre, 78180 Montigny le Bretonneux, France.
Background:
Remifentanil and propofol are often used in combination for general anesthesia. We developed a method using LC-MS for their simultaneous measurement in human plasma. Methodology & results: After addition of remifentanil-(13)C6 and propofol-d18 (IS), 500 µl of plasma were extracted with ethylacetate/hexane. Analysis conditions included gradient elution (water/acetonitrile), electrospray ionization and detection with a triple quadripole mass spectrometer. Remifentanil and propofol were ionized in the positive and negative mode, respectively. The method was validated according to the European Medicines Agency guideline for the validation of bioanalytical methods, then successfully applied to clinical samples from three patients who had undergone liver transplantation.
Conclusion:
This method is suitable for the simultaneous quantification of remifentanil and propofol in clinical studies.
More Related Videos
10:17High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
08:56Detection of Regulated Ergot Alkaloids in Food Matrices by Liquid Chromatography-Trapped Ion Mobility Spectrometry-Time-of-Flight Mass Spectrometry
Published on: November 22, 2024