Related Experiment Video
Updated: Apr 6, 2026

Analyzing Long-Term Electrocardiography Recordings to Detect Arrhythmias in Mice
Published on: May 23, 2021
Heartbeat Cycle Length Detection by a Ballistocardiographic Sensor in Atrial Fibrillation and Sinus Rhythm
Matthias Daniel Zink1, Christoph Brüser2, Patrick Winnersbach1
1Department of Cardiology, Pneumology, Angiology and Intensive Care Medicine, University Hospital, RWTH Aachen University, Pauwelsstrasse 30, 52074 Aachen, Germany.
Background:
Heart rate monitoring is especially interesting in patients with atrial fibrillation (AF) and is routinely performed by ECG. A ballistocardiography (BCG) foil is an unobtrusive sensor for mechanical vibrations. We tested the correlation of heartbeat cycle length detection by a novel algorithm for a BCG foil to an ECG in AF and sinus rhythm (SR).
Methods:
In 22 patients we obtained BCG and synchronized ECG recordings before and after cardioversion and examined the correlation between heartbeat characteristics.
Results:
We analyzed a total of 4317 heartbeats during AF and 2445 during SR with a correlation between ECG and BCG during AF of r = 0.70 (95% CI 0.68-0.71, P < 0.0001) and r = 0.75 (95% CI 0.73-0.77, P < 0.0001) during SR. By adding a quality index, artifacts could be reduced and the correlation increased for AF to 0.76 (95% CI 0.74-0.77, P < 0.0001, n = 3468) and for SR to 0.85 (95% CI 0.83-0.86, P < 0.0001, n = 2176).
Conclusion:
Heartbeat cycle length measurement by our novel algorithm for BCG foil is feasible during SR and AF, offering new possibilities of unobtrusive heart rate monitoring. This trial is registered with IRB registration number EK205/11. This trial is registered with clinical trials registration number NCT01779674.
More Related Videos
08:10Estimating Bilateral Atrial Function by Cardiovascular Magnetic Resonance Feature Tracking in Patients with Paroxysmal Atrial Fibrillation
Published on: July 20, 2022
09:17High-Resolution Endocardial and Epicardial Optical Mapping in a Sheep Model of Stretch-Induced Atrial Fibrillation
Published on: July 29, 2011
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Dysrhythmias IV: Characteristics of Bradyarrhythmias
ECG Interpretation of Arrhythmias I: Sinus Arrhythmias
Types of Arrhythmias
Sinus Node Arrhythmias
Sinus Bradycardia: Originating from the sinoatrial (SA) node, sinus bradycardia involves slower impulses, resulting in a heart rate of less than 60 beats per minute (bpm). Causes include sleep, vagal stimulation, beta-blockers, hypothyroidism,...
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...
Dysrhythmias III: Characteristics of Dysrhythmias