Radiation reduction and faster acquisition times with SPECT gated blood pool scans using a high-efficiency cardiac
W Lane Duvall1, Krista A Guma-Demers2, Titus George2
1Hartford Hospital Division of Cardiology (Henry Low Heart Center), Hartford Hospital, 80 Seymour Street, Hartford, CT, 06102, USA. lane.duvall@hhchealth.org.
Background:
Planar gated blood pool scans are an established method for the evaluation of left ventricular ejection fraction (LVEF) but the camera technology used for these studies has not significantly changed in decades. The purpose of this study was to determine the diagnostic accuracy of new high-efficiency SPECT gated blood pool scans compared to traditional scans and determine if they can be performed with lower radiation doses or faster acquisition times.
Methods:
Patients undergoing a planar gated blood pool scan on a Na-I SPECT camera who consented to participate were subsequently imaged for 5 minutes in "List Mode" using a high-efficiency SPECT camera. LVEF was calculated for both the planar study and at 1, 2, 3, 4, and 5 minutes of acquisition on the high-efficiency camera. Counts acquired in the field of view, counts in the cardiac blood pool and LVEF were compared.
Results:
A total of 46 patients were analyzed (48% male, mean age 55 years, and BMI 27.6 kg/m(2)) who received an average Tc-99m dose of 20.3 mCi (5.3 mSv), 17 (37%) with abnormal LVEF's. The Na-I camera averaged 24,514 counts/min/mCi in the field of view and 8662 counts/min/mCi in the cardiac blood pool while the high-efficiency camera averaged 65,219 counts/min/mCi and 41,427 counts/min/mCi, respectively. Compared to the planar calculation of LVEF, 1-minute SPECT LVEF was on average 8.6 ± 10.7 higher, 2 minutes 3.5 ± 7.6 higher, 3 minutes 2.9 ± 8.5 higher, 4 minutes 2.5 ± 7.0 higher, and 5 minutes 1.1 ± 6.2 higher. Good correlation was seen between the SPECT LVEF's and the planar LVEF's across all acquisition times with correlation coefficients of 0.74-0.93.
Conclusions:
High-efficiency SPECT technology can reduce radiation exposure to patients during gated blood pool imaging or decrease acquisition time while maintaining diagnostic accuracy. Based on the improved count sensitivity with high-efficiency SPECT, a 50% reduction in injected activity may be achievable while maintaining short imaging times of 5 minutes, with further reduction possible at longer imaging times.
More Related Videos
11:09High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
Published on: December 16, 2022
10:21Continuous Blood Sampling in Small Animal Positron Emission Tomography/Computed Tomography Enables the Measurement of the Arterial Input Function
Published on: August 8, 2019
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...