A Plate-Based Assay for the Measurement of Endogenous Monoamine Release in Acute Brain Slices
Published on: August 11, 2021
3.9K
Quantification of Drugs in Brain Samples.
Birgit Reiter1, Thomas Stimpfl2
1Forensic Toxicology, Clinical Institute of Laboratory Medicine, Medical University of Vienna, Waehringer Guertel 18-20, Vienna A-1090, Austria.
Journal of Analytical Toxicology
|August 2, 2015
Summary
Quantifying substances in postmortem samples is difficult without certified reference materials. This study validates an automated solid-phase extraction method for reliable quantification using the standard addition technique, crucial for forensic toxicology.
Area of Science:
- Forensic Toxicology
- Analytical Chemistry
- Biochemistry
Background:
- Reliable quantification in complex postmortem matrices is challenging due to the lack of certified reference materials.
- Standard addition methods are viable but require highly reproducible extraction procedures.
- Automated solid-phase extraction (SPE) offers a potential solution for consistent sample preparation.
Purpose of the Study:
- To validate an automated solid-phase extraction (SPE) screening procedure for postmortem analyses.
- To assess the method's suitability for standard addition quantification in complex matrices.
- To evaluate the impact of initial concentration estimates on calibration curve accuracy.
Main Methods:
- Validation of an automated SPE method using fortified pig brain samples.
- Inclusion of nine representative compounds to simulate realistic postmortem conditions.
- Assessment of validation parameters: linearity, repeatability, reproducibility, and recovery.
Main Results:
- The automated SPE method demonstrated satisfactory performance across all tested validation parameters.
- The method proved suitable for standard addition experiments, despite not being optimized for the specific compounds.
- Accurate concentration estimation is critical for calibration curve reliability; underestimation is preferable to overestimation.
Conclusions:
- A validated automated SPE method can be effectively employed for standard addition quantification in postmortem samples.
- The accuracy of the initial concentration estimate significantly influences the reliability of the final quantification.
- This validated method facilitates further research into optimizing standard addition protocols in forensic analysis.