Cortical Bone Assessment Using Ultrasonic Guided Waves: A Reproducibility Study in a Healthy Population
Published on: January 31, 2025
Ultrasonic wave properties of human bone marrow in the femur and tibia
Satoshi Kawasaki1, Ryohei Ueda1, Akihiko Hasegawa2
1Laboratory of Ultrasonic Electronics, Doshisha University, 1-3 Tatara-Miyakodani, Kyotanabe, Kyoto 610-0321, Japan.
Abstract:
Ultrasonic wave properties of human bone marrow obtained in the femur and tibia were measured using an ultrasound pulse technique. The measured frequency range was 4-10 MHz, and the temperature range was 30 °C-40 °C. The sound velocity was 1410 m/s, and the attenuation coefficient was 4.4 dB/cm at 36 °C (10 MHz). These values decreased with temperature. Site dependence and individual differences in elderly human bone marrow were negligible. The slopes of the attenuation coefficient were estimated by a power law. The values of the exponent n were 2.0 (30 °C-38 °C) and 2.3 (40 °C).
Related Concept Videos
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Bones of the Lower Limb: Femur and Patella
The Bone Matrix
Compact Bone
Compact bone, also called cortical bone, is the denser, stronger of the two types of bone tissue. It is found under the periosteum and in the diaphyses of long bones, where it provides support and protection. The microscopic structural unit of compact bone is called an osteon, or haversian system. Each osteon is composed of concentric rings of calcified...