Articular congruency of the Salto Talaris total ankle prosthesis
Colin H Morris1, Jeffrey C Christensen2, Randal P Ching3
1University of Washington, Seattle, WA, United States.
Background:
The Salto-Talaris polyethylene articulating surface was designed to allow, but limit accessory motion. This investigation examines surface characteristics between the polyethylene bearing and anatomic talar component in various positions of function.
Methods:
A Salto Talaris talar prosthesis and matching polyethylene bearing were scanned to create digital solid body models and manipulated to assess surface contact during simulated gait. With computer micromanipulation of the component positions, the surface intersections were recorded for 15 different alignments.
Results:
The Salto Talaris has limited contact congruity with four points of contact in dorsiflexion, neutral, and plantarflexion. Lateral and medial translations showed only 2-point contact. The radii of curvatures between the talar component and polyethylene surfaces do not match. There was no sulcus contact yet component separation distance was small, suggesting increased loads.
Conclusion:
Surface incongruency was measured based on computer model analysis which raises a concern of increased contact pressures.