Operator radiation exposure during right or left transradial coronary angiography: A phantom study
Alessandro Sciahbasi1, Stefano Rigattieri1, Alessandro Sarandrea2
1Interventional Cardiology, Sandro Pertini Hospital, Rome, Italy.
Background:
Previous studies showed a possible lower radiation dose absorbed by operators comparing LRA and RRA for percutaneous coronary procedures. The reasons of this lower radiation dose are not well known. The aim of this study was to evaluate the radiation dose absorbed by operators comparing left with right radial access (LRA and RRA respectively) during a simulated diagnostic coronary angiography using a phantom.
Methods:
A coronary angiography examination was simulated on a phantom by 5 operators using eight projections with 5 seconds fluoroscopy each. Each operator was equipped with 4 electronic dosimeters placed at thorax, at left wrist, at left head and at hip level. Radiation doses were expressed in picosievert and normalized by dose area product.
Results:
LRA compared to RRA was associated with a significant lower operator dose at wrist (36pSv/cGYcm(2) [IQR 18-59pSv/cGYcm(2)] and 48pSv/cGYcm(2) [IQR 22-148pSv/cGYcm(2)] respectively, p=0.01) and thorax (3pSv/cGYcm(2) [IQR 2-5pSv/cGYcm(2)] and 10pSv/cGYcm(2) [6-23pSv/cGYcm(2)] respectively, p<0.001) but with a significant higher radiation dose at hip level (102pSv/cGYcm(2) [IQR 44-199pSv/cGYcm(2)] and 67pSv/cGYcm(2) [IQR 39-132pSv/cGYcm(2)] respectively, p=0.02). Conversely the radiation dose at left side of the head did not show significant differences between the two approaches.
Conclusions:
In this phantom study simulating a diagnostic coronarography the use of LRA compared to RRA was associated with a significant lower radiation dose at wrist and thorax but with an increased dose at hip level.
Summary:
To evaluate the radiation dose absorbed by operators comparing left with right radial access (LRA and RRA respectively) we simulated a diagnostic coronary angiography using a dedicated phantom. Operators were equipped with dedicated electronic dosimeters at wrist, hip, head and thorax level. LRA compared to RRA was associated with a significant lower operator dose at wrist and thorax but with a significant higher radiation dose at hip level whereas the radiation dose at left side of the head did not show significant differences between the two approaches.
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System V: CT
X-ray Imaging
Imaging Studies VII: Vascular Imaging
Radiological Investigation I: X-ray and CT