Related Experiment Video
Updated: Apr 6, 2026

Laparoscopic-Assisted Seldinger Technique for Peritoneal Dialysis Catheter Insertion
Published on: May 23, 2025
Initial Clinical Experience: Symmetric-Tip Dialysis Catheter with Helical Flow Characteristics Improves Patient
Timothy W I Clark1, Jonas W Redmond1, Mark P Mantell2
1Section of Interventional Radiology, Department of Radiology Penn Presbyterian Medical Center, University of Pennsylvania, 39th and Market Streets, Philadelphia, PA 19104.
Purpose:
To report preliminary clinical experience with a new symmetric-tip dialysis catheter compared with a conventional split-tip catheter.
Materials And Methods:
Over a 5-month period, patients requiring a tunneled catheter for hemodialysis or undergoing exchange of a dysfunctional dialysis catheter at a tertiary academic medical center were retrospectively analyzed. Patients underwent placement of a VectorFlow or Ash Split Cath catheter at the discretion of the inserting interventional radiologist. Patient demographics, catheter patency, mean blood flow rate, and arterial and venous pressures were compared according to catheter type. Catheter failure was analyzed based on clinical and anatomic variables by using a multivariate Cox proportional-hazards model.
Results:
A total of 33 VectorFlow and 46 Ash Split Cath catheters were placed. Patients in the VectorFlow group had significantly higher body mass index (P = .013) and Charlson Comorbidity Index (P = .049), as well as more non-internal jugular vein placements. At 120 days, 89% of VectorFlow catheters remained functional, compared with 45% of Ash Split Cath catheters (P = .046). The VectorFlow catheter was associated with 16% lower arterial pressures during dialysis (P = .009); mean blood flow rate was equivalent. On multivariate analysis, the risk of catheter failure was 13.3 times higher in the Ash Split Cath group compared with the VectorFlow group (P = .004). Left-sided catheters were also predictive of catheter failure (relative risk = 5.5; P = .02).
Conclusions:
The VectorFlow catheter was associated with a significant increase in intervention-free catheter patency compared with the Ash Split Cath catheter, with equivalent flow at lower arterial pressures during dialysis.
Related Concept Videos
Hemodialysis I: Introduction
Hemodialysis II: Procedure and Complications
Peritoneal Dialysis I: Introduction and Procedure
Hemodialysis III: Nursing Management
Dialysis
Dialysis
Acute kidney injury develops suddenly and can be caused by pre-renal causes (e.g., hypovolemia, shock), intrinsic renal causes (e.g., acute tubular necrosis), or post-renal causes (e.g., urinary obstruction). In contrast, chronic renal failure progresses gradually over time and is often...