The impact of pediatric-specific dose modulation curves on radiation dose and image quality in head computed
Joana Santos1, Shane Foley2, Graciano Paulo3
1Instituto Politécnico de Coimbra, ESTESC, DMIR, Rua 5 de Outubro, S. Martinho do Bispo, 3046-854, Coimbra, Portugal. joanasantos@estescoimbra.pt.
Background:
The volume of CT examinations has increased with resultant increases in collective dose values over the last decade.
Objective:
To analyze the impact of the tube current and voltage modulation for dose values and image quality of pediatric head CT examinations.
Materials And Methods:
Head CT examinations were performed on anthropomorphic phantoms and four pediatric age categories before and after the introduction of dedicated pediatric curves for tube voltage and current modulation. Local diagnostic reference levels were calculated. Visual grading characteristic image quality evaluation was performed by four pediatric neuroradiologists and image noise comparisons were performed.
Results:
Pediatric-specific modulation curves demonstrated a 49% decrease in mean radiation dose for phantom examinations. The local diagnostic reference levels (CTDIvol) for clinical examinations decreased by 52%, 41%, 46% and 40% for newborn, 5-, 10- and 15-year-old patients, respectively. Visual grading characteristic image quality was maintained for the majority of age categorizations (area under the curve = 0.5) and image noise measurements did not change (P = 0.693).
Conclusion:
Pediatric-specific dose modulation curves resulted in an overall mean dose reduction of 45% with no significant differences in subjective or objective image quality findings.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Dose Response Curve: Conventional Versus Nonmonotonic
Imaging Studies III: Computed Tomography
Drug Dosing: Infants and Children