Comparison of Agreement and Accuracy using Binocular Wavefront Optometer with Autorefractor and Phoropter
Published on: September 16, 2025
Comparing Autorefractors for Measurement of Accommodation
Mikel Aldaba1, Selena Gómez-López, Meritxell Vilaseca
1*PhD †MSc Centre for Sensors, Instruments, and Systems Development (CD6), Universitat Politecnica de Catalunya (UPC), Terrassa, Barcelona, Spain (all authors).
The WAM-5500 autorefractor showed good agreement with subjective refraction, while the PowerRef-II autorefractor yielded more hyperopic results. Dynamic measurements revealed the PowerRef-II captured faster accommodative velocities, influenced by its higher sampling frequency.
Area of Science:
- Ophthalmology
- Optometry
- Vision Science
Background:
- Accurate measurement of accommodative response is crucial for diagnosing and managing refractive errors and binocular vision anomalies.
- Autorefractors are widely used tools for objective refraction, but their performance in measuring dynamic and static accommodation can vary.
- Understanding the differences between autorefractors like the WAM-5500 and PowerRef-II is important for clinical decision-making.
Purpose of the Study:
- To compare the static and dynamic accommodative responses measured by the WAM-5500 and PowerRef-II autorefractors.
- To evaluate the agreement between subjective refraction and objective measurements from both instruments.
Main Methods:
- Monocular static and dynamic accommodative responses were measured in 30 pre-presbyopic individuals using the WAM-5500 and PowerRef-II.
- Static measurements involved spherical equivalent determination at 0.00, 2.50, and 5.00 diopters (D) of accommodative stimulus.
- Dynamic measurements assessed accommodation and disaccommodation velocities during vergence changes, with PowerRef-II data analyzed at 5 Hz and 25 Hz sampling frequencies.
Main Results:
- Static measurements showed good agreement between subjective refraction and WAM-5500 (0.07 ±0.21 D), but less agreement with PowerRef-II (0.70 ±0.47 D) at distance.
- Differences between instruments were not significant at 2.50 D static stimulation but were significant at 5.00 D (-0.32 ±0.48 D).
- The PowerRef-II, particularly at 25 Hz, recorded significantly faster mean and peak velocities for accommodation and disaccommodation compared to the WAM-5500.
Conclusions:
- The WAM-5500 demonstrates good agreement with subjective refraction for static accommodation.
- The PowerRef-II tends to produce more hyperopic static results and captures faster dynamic accommodative responses, likely due to its higher sampling frequency.
- Clinical use of the PowerRef-II should consider its higher sensitivity to dynamic changes and potential for more hyperopic static readings.