Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

Bones of the Lower Limb: Femur and Patella01:16

Bones of the Lower Limb: Femur and Patella

9.3K
The femur is the body's longest and strongest bone spanning the thigh region. Its head articulates with the acetabulum of the hip bone to form the hip joint. A minor indentation on the medial side of the femoral head, called the fovea capitis, serves as the site of attachment for the ligament of the head of the femur. This weak ligament spans the femur and acetabulum and supports the hip joint. The narrowed region below the head is the neck of the femur. The inclination angle between the...
9.3K
Wedges01:24

Wedges

2.0K
A wedge is a simple machine that serves various purposes, such as adjusting the elevation of structural or mechanical parts, providing stability for heavy objects, and splitting a body into two parts. This versatile tool can amplify an applied force, making it easier to manipulate large or heavy objects.
Consider using a wedge to lift a heavy slab. Here, the wedge functions by converting the applied force into a much larger force directed almost perpendicular to the initial force. This...
2.0K
Self-Locking Screw01:16

Self-Locking Screw

2.7K
A square-threaded screw jack is a mechanical device widely used for lifting heavy loads or applying considerable force. One of the key features that can make a screw jack more effective and reliable is its self-locking capability.
A square-threaded screw jack carrying a load is considered self-locking if the screw retains its position even after the moment applied to it is removed.
2.7K

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Use of an Artificial Intelligence Device for Evaluating Blood Loss in Complex Major Orthopaedic Surgery Procedures.

The Journal of arthroplasty·2024
Same author

Clinical Results of Revision TKA in Patients With Presumed Metal and Cement Allergy.

The Journal of arthroplasty·2022
Same author

Periacetabular Osteotomy in United States Military Personnel: Results From an Interservice Hip Preservation Practice.

Orthopaedic journal of sports medicine·2022
Same author

Cobalt-chromium femoral components developed scratches and released metal debris in simulated wear whereas ceramic femoral components did not.

The bone & joint journal·2021
Same author

Acetabular labral tears in the adolescent athlete: results of a graduated management protocol from therapy to arthroscopy.

Journal of pediatric orthopedics. Part B·2020
Same author

Developmental Dysplasia of the Hip in Adolescents and Young Adults.

The Journal of the American Academy of Orthopaedic Surgeons·2020

Related Experiment Video

Updated: Apr 5, 2026

The Transition to an Anterior-Based Muscle Sparing Approach Improves Early Postoperative Function but is Associated with a Learning Curve
09:51

The Transition to an Anterior-Based Muscle Sparing Approach Improves Early Postoperative Function but is Associated with a Learning Curve

Published on: September 7, 2022

3.8K

Thin-Walled Cross-Linked Acetabular Liners Need Not Exhibit Reduced Locking Strength.

Andrew S Murtha, Marcel E Roy, Leo A Whiteside

    Orthopedics
    |August 14, 2015
    PubMed
    Summary

    This study tested whether thinning the walls of hip replacement liners affects their locking strength when using larger femoral heads. Researchers used lever-out and torsion tests on three liner sizes made of cross-linked polyethylene. They found that larger heads did not reduce the locking strength of thinned liners. The results suggest that these liners remain biomechanically sound even when thinned to at least 2.86 mm. The study focused on a specific liner design called MicroSeal. The findings indicate that using larger femoral heads with thinned liners is a viable option in hip replacement surgery.

    Keywords:
    hip arthroplastyacetabular liner designbiomechanical testingfemoral head size

    Frequently Asked Questions

    More Related Videos

    Treatment with Locking Intramedullary Nailing for Intertrochanteric Fracture of the Femur Utilizing a New Awl with a Distal Positioner
    04:41

    Treatment with Locking Intramedullary Nailing for Intertrochanteric Fracture of the Femur Utilizing a New Awl with a Distal Positioner

    Published on: June 6, 2025

    1.5K
    Author Spotlight: A Novel 3D-Printed Titanium Implant for Minimally Invasive Treatment of Hip Dysplasia in Young Dogs
    08:40

    Author Spotlight: A Novel 3D-Printed Titanium Implant for Minimally Invasive Treatment of Hip Dysplasia in Young Dogs

    Published on: April 19, 2024

    4.0K

    Related Experiment Videos

    Last Updated: Apr 5, 2026

    The Transition to an Anterior-Based Muscle Sparing Approach Improves Early Postoperative Function but is Associated with a Learning Curve
    09:51

    The Transition to an Anterior-Based Muscle Sparing Approach Improves Early Postoperative Function but is Associated with a Learning Curve

    Published on: September 7, 2022

    3.8K
    Treatment with Locking Intramedullary Nailing for Intertrochanteric Fracture of the Femur Utilizing a New Awl with a Distal Positioner
    04:41

    Treatment with Locking Intramedullary Nailing for Intertrochanteric Fracture of the Femur Utilizing a New Awl with a Distal Positioner

    Published on: June 6, 2025

    1.5K
    Author Spotlight: A Novel 3D-Printed Titanium Implant for Minimally Invasive Treatment of Hip Dysplasia in Young Dogs
    08:40

    Author Spotlight: A Novel 3D-Printed Titanium Implant for Minimally Invasive Treatment of Hip Dysplasia in Young Dogs

    Published on: April 19, 2024

    4.0K

    Area of Science:

    • Orthopedic surgery outcomes research within biomechanics
    • Medical device engineering in joint replacement

    Background:

    Current knowledge suggests that larger femoral heads in hip replacements may lower dislocation risks. However, the mechanical effects of reducing liner wall thickness remain unclear. Prior research has shown that cross-linked polyethylene improves wear resistance. But no prior work had resolved how thinning these liners affects locking strength. This uncertainty drove the need to assess mechanical integrity in thinned liners. Lever-out and torsion tests are standard methods for evaluating locking mechanisms. Yet, the impact of varying head sizes on these mechanisms is not fully understood. The study aimed to address this gap by testing specific liner sizes. The findings could clarify whether thinner liners maintain sufficient strength for clinical use.

    Purpose Of The Study:

    The study aimed to evaluate the mechanical integrity of locking mechanisms in thinned cross-linked polyethylene acetabular liners. Larger femoral heads require thinner walls, which may affect locking strength. The researchers wanted to determine if this thinning compromises the liner's stability. They focused on three specific liner sizes with different head and neck dimensions. The goal was to assess failure strength using lever-out and torsion tests. The study sought to confirm whether reduced wall thickness impacts locking performance. By isolating locking tabs from central post contributions, the tests aimed to be precise. The results could inform whether these liners remain biomechanically sound for clinical use.

    Main Methods:

    Researchers used lever-out and torsion tests to assess the locking mechanisms of acetabular liners. The liners were made of cross-linked and re-melted UHMWPE in three sizes: 50/28, 50/36, and 52/36 mm. Lever-out tests measured the force needed to dislodge the liner from its shell. Torsion tests were conducted at 45° and 90° angles to evaluate rotational strength. These angles helped isolate the locking tabs from the central post's influence. The study compared failure strengths across the different liner sizes. Statistical analysis determined whether differences were significant. The methods allowed a focused evaluation of the liner's mechanical behavior.

    Main Results:

    Lever-out tests showed a trend toward reduced failure strength in 50/36-mm liners compared to 50/28-mm ones. The 50/36-mm liners had a failure strength of 13.3 N · m versus 12.3 N · m for 50/28-mm. The difference was not statistically significant (P = 0.0502). The 52/36-mm liners showed a lever-out strength of 12.2 ± 0.94 N · m. Torsion tests revealed similar failure torques between 50/28- and 50/36-mm liners at both angles. However, 52/36-mm liners had significantly higher failure torques at each angle. The results suggest that larger femoral heads do not reduce locking strength. The findings indicate that thinned liners maintain sufficient mechanical integrity.

    Conclusions:

    The authors propose that larger femoral heads do not compromise the locking mechanism of thinned liners. The study suggests that liners with at least 2.86 mm thickness remain biomechanically sound. The results indicate that lever-out and torsion strengths are not significantly reduced in thinner liners. The findings support the use of cross-linked UHMWPE liners with larger heads. The study does not claim that all liner designs behave the same way. The specific design tested, MicroSeal, appears to maintain locking strength when thinned. The authors do not generalize these results to other liner types or manufacturers. The conclusions are limited to the tested sizes and materials.

    According to the authors, larger femoral heads do not compromise the locking strength of thinned MicroSeal liners.

    The study tested liners with a minimum wall thickness of 2.86 mm.

    Torsion tests at 45° and 90° isolated the locking tabs from the central post's influence.

    Failure torque in 52/36-mm liners was significantly higher than in 50/28- and 50/36-mm liners.

    Lever-out strength in 50/36-mm liners was 13.3 N · m.

    The study suggests that cross-linked UHMWPE liners with larger heads remain biomechanically sound when thinned.