Quantitative precision of optical frequency domain imaging: direct comparison with frequency domain optical coherence
Yuhei Kobayashi1, Hideki Kitahara1, Shigemitsu Tanaka1
1Division of Cardiovascular Medicine, Stanford Cardiovascular Institute, Stanford University School of Medicine, 300 Pasteur Drive, Room H3554, Stanford, CA, 94305-5637, USA.
Abstract:
No systematic validation study is available with optical frequency domain imaging (OFDI), directly compared with frequency domain optical coherence tomography (FD-OCT) and intravascular ultrasound (IVUS). Controversy also remains about the impact of different stent contour tracing methods by OFDI/FD-OCT. In vitro: coronary phantom models (1.51-5.04 mm) were imaged with OFDI, FD-OCT, and IVUS, demonstrating excellent quantitative precision with a slight overestimation of mean lumen diameter (difference 0.01-0.02 mm). In vivo: corresponding 64 OFDI/IVUS images of stented coronary segments from 20 swines were analyzed. Minimum lumen area by OFDI was larger than IVUS at baseline (P < 0.001), whereas it was smaller than IVUS at follow-up. When stent was traced at leading edges of struts by OFDI, minimum stent area was similar between OFDI and IVUS (P = 0.60). When traced at the highest intensity points of struts by OFDI, it was significantly larger in OFDI than in IVUS (P < 0.001). Three modalities have clinically acceptable precision across the wide range of lumen diameters. In vivo measurements by OFDI and IVUS could slightly be discrepant depending on the parameters and time points. In stent assessment by OFDI, the 2 methods led to a small but systematic difference; therefore, consistency in methodology is advised for comparative studies.
More Related Videos
14:09Quantitative Optical Microscopy: Measurement of Cellular Biophysical Features with a Standard Optical Microscope
Published on: April 7, 2014
14:21Optical Frequency Domain Imaging of Ex vivo Pulmonary Resection Specimens: Obtaining One to One Image to Histopathology Correlation
Published on: January 22, 2013
Related Concept Videos
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
Uncertainty in Measurement: Accuracy and Precision
Accuracy and Precision
Uncertainty in Measurement: Reading Instruments