The Contribution of Common Surgically Implanted Hardware to Functional MR Imaging Artifacts
A A Desai1, M K Strother2, C C Faraco2
1From the Departments of Radiology and Radiological Sciences (A.A.D., M.K.S., C.C.F., V.L.M., T.R.L., L.M.D., M.J.D.) aditi.desai@vanderbilt.edu.
AJNR. American Journal of Neuroradiology
|August 15, 2015
Summary
Metallic implants cause artifacts in blood oxygenation level-dependent (BOLD) MRI scans. While most intracranial hardware causes minor issues, extracranial hardware and ventriculoperitoneal shunts can significantly limit functional imaging interpretation.
Area of Science:
- Medical Imaging
- Neuroimaging
- Radiology
Background:
- Blood oxygenation level-dependent (BOLD) MRI is crucial for assessing brain function.
- Metallic implants can cause susceptibility artifacts, compromising BOLD MRI data.
- Assessing the impact of surgical hardware on BOLD MRI is essential for clinical interpretation.
Purpose of the Study:
- To quantify the volume of BOLD MRI artifacts caused by metallic implants.
- To evaluate the impact of these artifacts on the clinical interpretability of functional MRI data.
- To identify specific types of hardware that pose the greatest challenge to BOLD MRI.
Main Methods:
- Retrospective analysis of 343 BOLD MRIs (3T, gradient echo EPI) acquired between 2006-2014.
- Categorization of patients by cerebrovascular disease or functional localization (language/motor).
- Measurement of artifact volume and assessment of interpretability by a neuroradiologist.
Main Results:
- Intracranial hardware produced mean artifact volumes of 4.3 cm³ with no-to-mild interpretability impact.
- Extracranial hardware resulted in larger artifacts (28.4-39.9 cm³), with no-to-moderate impact, except for one case.
- Ventriculoperitoneal shunts generated the largest artifacts (mean 95.7 cm³), causing moderate-to-severe limitations in interpretation.
Conclusions:
- BOLD MRI provides interpretable functional maps for most patients with metallic implants, even with artifacts up to 30-40 cm³.
- Ventriculoperitoneal shunts, especially those with programmable valves, and numerous KLS-Martin maxDrive screws are exceptions.
- Careful consideration of hardware type and location is necessary for accurate BOLD MRI interpretation.
Related Concept Videos
Magnetic Resonance Imaging
10.4K
Magnetic resonance imaging (MRI) is a noninvasive medical imaging technique based on a phenomenon of nuclear physics discovered in the 1930s, in which matter exposed to magnetic fields and radio waves was found to emit radio signals. In 1970, a physician and researcher named Raymond Damadian noticed that malignant (cancerous) tissue gave off different signals than normal body tissue. He applied for a patent for the first MRI scanning device in clinical use by the early 1980s. The early MRI...
10.4K
Imaging Studies for Cardiovascular System IV: CMRI
542
Cardiovascular magnetic resonance imaging, or CMRI, is a non-invasive diagnostic test that employs a magnetic field and radiofrequency waves to create precise images of the heart and arteries. It provides comprehensive information about cardiac anatomy, function, perfusion, and tissue characterization without ionizing radiation.IndicationsCMRI diagnoses various heart conditions, including tissue damage from heart attacks, ischemic heart disease, myocarditis, aortic issues (tears, aneurysms,...
542
Imaging Studies I: CT and MRI
1.2K
Introduction: MRI and CT scans are crucial advancements in medical imaging techniques, playing a vital role in diagnosing conditions related to the gastrointestinal (GI) system. Each scan serves distinct purposes, targets specific areas, and requires unique nursing duties.
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
1.2K