Assessment of Magnetic Resonance Imaging Artifact Following Cervical Total Disc Arthroplasty
Amir H Fayyazi1, Jennifer Taormina2, David Svach3
1OAA Orthopaedic Specialists, Allentown, PA.
Background:
Cervical disc arthroplasty has become a technique for the treatment of cervical degenerative disc disease. Clinically, the need to accurately assess the neural elements at the operative and adjacent levels is critical postoperatively. The purpose of this study was to quantitatively and qualitatively measure the amount of MRI artifact produced by various cervical total disc replacements.
Methods:
T1 and T2-weighted turbo spin-echo MRI sequences were collected on the cervical spine (C2-T1) of a 68 year-old unembalmed male cadaver. A discectomy was performed at C5-6, followed by successive implantation of six different total disc replacements. The scans were quantitatively evaluated by three of the authors. The volume of artifact was measured using image analysis software. Qualitative analysis of the adjacent and index neural elements was performed.
Results:
The artifact in the T2 weighted images was noted to be 58.6±7.3 cm3 for Prestige ST, 14.2±1.3 cm3 for ProDisc-C, 7.5±0.8 cm3 for Discover, 8.0±0.3 cm3 for Prestige LP, 6.6±0.7 cm3 for Bryan, and 7.3±0.6 cm3 for ProDisc-C titanium prototype. Acceptable intraobserver and excellent interobsever correlation was demonstrated using Pearson Correlation and Concordance Correlation Coefficient analysis. The adjacent and implanted level neural elements (spinal cord and neuroforamina) were easily visualized on the T2 weighted images after the implantation of titanium devices (ProDisc-C titanium prototype, Discover, Prestige LP and Bryan). After implantation of a cobalt chrome implant (ProDisc-C), the adjacent level neural elements were easily visualized but the implanted level could not be fully visualized due to distortion of the images. The quality of the distortion was least favorable after the implantation of the stainless steel implant (Prestige ST), where neither the adjacent nor the index level could be fully visualized.
Conclusion:
The volume of the artifact seen following cervical total disc arthroplasty is highly dependent upon the material property of the implant. Quantitative analysis described in this study demonstrated sufficiently low intraobserver and interobserver variability to be considered a reliable technique.
More Related Videos
07:06Surgical Technique for the Implantation of a Biomimetic Artificial Intervertebral Disc in a Goat Animal Model
Published on: October 10, 2025
03:24Author Spotlight: Development and Application of a Novel Suture Technique for Annular Fibrosus Repair in Percutaneous Transforaminal Endoscopic Discectomy
Published on: January 26, 2024
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System IV: CMRI