Measurement of spatial distribution of neutrons and gamma rays for BNCT using multi-imaging plate system
Kenichi Tanaka1, Yoshinori Sakurai2, Hiroki Tanaka2
1Graduate School of Engineering, Hiroshima University, Higashi-Hiroshima, Japan.
Abstract:
Quality assurance of the spatial distributions of neutrons and gamma rays was tried using imaging plates (IPs) and converters to enhance the beam components in the epithermal neutron mode of the Kyoto University Reactor. The converters used were 4mm thick epoxy resin with B4C at 6.85 weight-percent (wt%) (10)B for epithermal neutrons, and 3mm thick carbon for gamma rays. Results suggested that the IP signal does not need a sensitivity correction regardless of the incident radiation that produces it.