Development of a low-dose protocol for cone beam CT examinations of the anterior maxilla in children
Jose A Hidalgo Rivas1, Keith Horner1, Badri Thiruvenkatachari1
11 School of Dentistry, The University of Manchester, Manchester Academic Health Science Centre, Manchester, UK.
Objective:
The aim of this study was to develop a low-dose protocol suitable for cone beam CT (CBCT) examination of an impacted maxillary canine in children by using a combination of dosimetry with subjective and objective image quality assessment.
Methods:
Radiation dose and image quality measurements were made using a dental CBCT machine. An image quality phantom was used to investigate the relationship between objective image quality and dose-area product (DAP) for a broad range of exposure settings. Subjective image quality assessment was achieved using a paediatric skull phantom submerged in a water bath for the same range of exposure settings. Eight clinicians assessed each CBCT data set for nine aspects of image quality using a five-point rating scale of agreement.
Results:
Acceptable image quality, defined using subjective judgements by the clinicians of the skull phantom images, was achievable with DAP values of 127 mGy cm(2) or greater and a polytetrafluoroethylene (PTFE) contrast-to-noise ratio (CNR) of 4.8 or greater. A cautious choice was made to recommend a low-dose protocol of 80 kV and 3 mA for implementation into clinical practice, corresponding to a DAP value of 146 mGy cm(2) and a PTFE CNR of 5.0.
Conclusion:
A low-dose protocol for this particular CBCT machine was established which represents as much as a 50% reduction compared with manufacturer's recommendations.
Advances In Knowledge:
To the authors' best knowledge, this is the first study that addresses dose optimization in paediatric clinical protocols in dental CBCT. Furthermore, this study explores the relationship between radiation dose, objective and subjective image quality.
More Related Videos
05:49Author Spotlight: Advancing CBCT and Digital Dental Image Integration with AI-Assisted Digitization
Published on: February 23, 2024
10:23Author Spotlight: Three-Dimensional Cephalometric Landmark Annotation Demonstration on Human Cone Beam Computed Tomography Scans
Published on: September 8, 2023
Related Concept Videos
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT
Imaging Studies III: Computed Tomography