TWO FACTORS INFLUENCING DOSE RECONSTRUCTION IN LOW DOSE RANGE: THE VARIABILITY OF BKG INTENSITY ON ONE INDIVIDUAL AND
Tengda Zhang1, Wenyi Zhang1, Zhixin Zhao1
1Institute of Radiation Medicine Chinese Academy of Medical Sciences, Baidi Street 238, Tianjin, China.
Abstract:
A fast and accurate retrospective dosimetry method for the triage is very important in radiation accidents. Electron paramagnetic resonance (EPR) fingernail dosimetry is a promising way to estimate radiation dose. This article presents two factors influencing dose reconstruction in low dose range: the variability of background signal (BKG) intensity on one individual and water content. Comparing the EPR spectrum of dried and humidified fingernail samples, it is necessary to add a procedure of dehydration before EPR measurements, so as to eliminate the deviation caused by water content. Besides, the BKGs of different fingers' nails are not the same as researchers thought previously, and the difference between maximum and minimum BKG intensities of one individual can reach 55.89 %. Meanwhile, the variability of the BKG intensity among individuals is large enough to impact precise dose reconstruction. Water within fingernails and instability of BKG are two reasons that cause the inaccuracy of radiation dose reconstruction in low-dosage level.