Transference of CALIPER pediatric reference intervals to biochemical assays on the Roche cobas 6000 and the Roche

Victoria Higgins1, Man Khun Chan2, Michelle Nieuwesteeg2

  • 1CALIPER Program, Pediatric Laboratory Medicine, The Hospital for Sick Children, Canada; Department of Laboratory Medicine & Pathobiology, University of Toronto, Toronto, ON, Canada.

Clinical Biochemistry
|August 23, 2015
PubMed

Insights

Pediatric reference intervals from the Canadian Laboratory Initiative on Pediatric Reference Intervals (CALIPER) were successfully transferred to Roche systems. This broadens the CALIPER database for harmonized pediatric laboratory testing.

Area of Science:

  • Clinical Chemistry
  • Pediatric Laboratory Medicine
  • Reference Interval Development

Background:

  • The Canadian Laboratory Initiative on Pediatric Reference Intervals (CALIPER) established pediatric reference intervals for biochemical markers on the Abbott Architect system.
  • Previous work successfully transferred CALIPER intervals to Roche, Beckman, Ortho, and Siemens assays.

Purpose of the Study:

  • To expand the CALIPER database by transferring and verifying pediatric reference intervals for 52 biochemical assays on Roche cobas 6000 and Roche Modular P systems.
  • To assess the transferability and verification of established pediatric reference intervals across different analytical platforms.

Main Methods:

  • Utilized Clinical and Laboratory Standards Institute (CLSI) guidelines (C28-A3 and EP9-A2) for reference interval transference.
  • Attempted transference for 16 assays on the Roche cobas 6000 and 36 assays on the Roche Modular P.
  • Verified transferable intervals using 100 healthy pediatric samples from the CALIPER cohort.

Main Results:

  • High correlation observed between assay systems, with 81% (cobas 6000) and 86% (Modular P) of assays showing transferable intervals from Abbott.
  • Bicarbonate and magnesium were not transferable to either Roche system; calcium and prealbumin were not transferable to the Modular P.
  • Verified transferability rates were 62% for cobas 6000 and 61% for Modular P.

Conclusions:

  • This study successfully extended the CALIPER database to include Roche cobas 6000 and Modular P analytical systems.
  • Facilitates wider application of CALIPER pediatric reference intervals in centers using Roche biochemical assays.
  • Supports future efforts to develop common reference intervals across multiple platforms for harmonized laboratory test interpretation in pediatric diagnostics.