Motion analysis of cardiopulmonary resuscitation
Nathalie Fournier1, Yves Godio-Raboutet2, Maxime Llari2
1SAMU 13 (EMS 13), RUSH Departement (Reanimation Urgence SAMU Hyperbarie), Hôpital de la Timone-(La Timone teaching hospital), 264 rue Saint-Pierre, Marseille 13005 France.
Objective:
Some cardiopulmonary resuscitation (CPR) monitoring devices were released in recent years. Some of them are motion sensors. There are no guidelines were to position future or present sensors during CPR. We evaluate the possible influence of the location of motion sensors by a high-speed camera during a CPR on a manikin.
Material And Methods:
We performed a motion analysis by a high-speed camera during chest compression (CC) on a manikin to quantify chest inhomogeneous displacements and rescuer motion.
Results:
Midline chest was found to have an inhomogeneous depth during CC (19 mm for the upper sternum, 27 mm for the middle of the sternum, and 47 mm for the xiphoid). Rescuer anatomy has a complex motion.
Conclusion:
The direct application of the sensor under the hand performing CC seems to be the more accurate solution if the device allows it.
Related Concept Videos
Cardiopulmonary Resuscitation I: Adult
Cardiopulmonary Resuscitation II: ACLS Airway Management
Cardiopulmonary Resuscitation III: AED Use
Cardiopulmonary Resuscitation IV: Pharmacological Management
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Assessment of Airway, Skin Color, and Use of Accessory Muscles
Introduction
The initial evaluation of a patient's respiratory system...