Related Experiment Video
Updated: Apr 5, 2026

'Bioluminescent' Reporter Phage for the Detection of Category A Bacterial Pathogens
Published on: July 8, 2011
Rapid detection of Bacillus anthracis by γ phage amplification and lateral flow immunochromatography
Christopher R Cox1, Kirk R Jensen1, Roy R Mondesire2
1Colorado School of Mines, Department of Chemistry and Geochemistry, Golden, CO 80401, United States.
Abstract:
New, rapid point-of-need diagnostic methods for Bacillus anthracis detection can enhance civil and military responses to accidental or deliberate dispersal of anthrax as a biological weapon. Current laboratory-based methods for clinical identification of B. anthracis require 12 to 120h, and are confirmed by plaque assay using the well-characterized γ typing phage, which requires an additional minimum of 24h for bacterial culture. To reduce testing time, the natural specificity of γ phage amplification was investigated in combination with lateral flow immunochromatography (LFI) for rapid, point-of-need B. anthracis detection. Phage-based LFI detection of B. anthracis Sterne was validated over a range of bacterial and phage concentrations with optimal detection achieved in as little as 2h from the onset of amplification with a threshold sensitivity of 2.5×10(4)cfu/mL. The novel use of γ phage amplification detected with a simple, inexpensive LFI assay provides a rapid, sensitive, highly accurate, and field-deployable method for diagnostic ID of B. anthracis in a fraction of the time required by conventional techniques, and without the need for extensive laboratory culture.