Resonance artefacts in modern pressure monitoring systems
1Department of Information Engineering, University of Florence, Via S. Marta 3, 50139, Florence, Italy. leonardo.bocchi@unifi.it.
Abstract:
Resonance in pressure monitoring catheters is a well-known problem which was studied several years ago. Current piezoelectric devices have mechanical properties providing a resonance frequency and damping factor that theoretically assure resonance-free data. However, in particular cases, the coupling between the device, the catheter, and the vascular compliance of the patient could introduce artefacts in clinical settings leading to wrong pressure waveforms and values displayed in the monitor. In this research work we study a laboratory model of a clinical setting to evaluate in which cases the compound (catheter and device) could cause resonances in an unacceptable range. The classical pop-test is expanded for analysing the effect of the catheter. Results indicate that the presence of different catheters may alter significantly the acquired signal, up to an unacceptable level. Particular care should be used in the selection of the appropriate catheter. In particular, smaller diameters introduce higher damping coefficient that could help in avoiding undesired oscillations.
Related Concept Videos
Double Resonance Techniques: Overview
Spin decoupling is usually achieved by...
Pressure Gauges
Bioreactor Controls-I
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Measurement of Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.