Accuracy and Injection Force of the Gla-300 Injection Device Compared With Other Commercialized Disposable Insulin
David Klonoff1, Irina Nayberg2, Marissa Thonius3
1Diabetes Research Institute, Mills-Peninsula Health Services, San Mateo, CA, USA dklonoff@diabetestechnology.org.
Background:
To deliver insulin glargine 300 U/mL (Gla-300), the widely used SoloSTAR(®) pen has been modified to allow for accurate and precise delivery of required insulin units in one-third of the volume compared with insulin glargine 100 U/mL, while improving usability. Here we compare the accuracy and injection force of 3 disposable insulin pens: Gla-300 SoloSTAR(®), FlexPen(®), and KwikPen™.
Methods:
For the accuracy assessment, 60 of each of the 3 tested devices were used for the delivery of 3 different doses (1 U, half-maximal dose, and maximal dose), which were measured gravimetrically. For the injection force assessment, 20 pens of each of the 3 types were tested twice at half-maximal and once at maximal dose, at an injection speed of 6 U/s.
Results:
All tested pens met the International Organization for Standardization (ISO) requirements for dosing accuracy, with Gla-300 SoloSTAR showing the lowest between-dose variation (greatest reproducibility) at all dose levels. Mean injection force was significantly lower for Gla-300 SoloSTAR than for the other 2 pens at both half maximal and maximal doses (P < .0271).
Conclusion:
All tested pens were accurate according to ISO criteria, and the Gla-300 SoloSTAR pen displayed the greatest reproducibility and lowest injection force of any of the 3 tested devices.
Related Concept Videos
Insulin Formulations: Types and Delivery
Short-acting insulins are divided into...
Insulin: Dosing Regimen and Adverse Effects
The basal dose constitutes about 40%-50% of the total daily dose, with the rest as premeal insulin. The mealtime insulin dose should mirror...
Parenteral Drug Delivery Systems: Injectables, Implants, and Infusion Devices
Oral Hypoglycemic Agents: Glinides
Glucagon-like Receptor Agonists
GLP-1, when administered in high doses intravenously, triggers insulin secretion, inhibits glucagon release, slows gastric emptying, reduces food intake, and restores normal insulin secretion. However, its rapid inactivation by...